FDA Medical devices recall ·
EOPA Elongated One-Piece Arterial Cannula DIL TIP 22FR… recall
- Recall number
- Z-2809-2026
- Date
- (initiated June 24, 2026)
- Company / brand
- Medtronic Perfusion Systems
7611 Northland Dr N, Brooklyn Park, MN, United States - Category
- Medical devices › Class II
- Classification
- Class II
- Status
- Ongoing
- Units affected
- 3,110 units
- Hazard tags
- Laceration
- Initiated by
- Voluntary: Firm initiated
Read the official FDA notice ↗
Product
EOPA Elongated One-Piece Arterial Cannula DIL TIP 22FR, intended for use with cardiopulmonary bypass as an arterial return cannula, Model Number 77722
Why was it recalled?
Pinhole leaks may occur in the wire-wound body during cardiopulmonary bypass. Potential risks associated with this condition may include hypovolemia, vessel laceration or perforation, hypotension, ischemia, and organ dysfunction.
What should you do?
FDA enforcement reports do not list a consumer remedy. Check the product codes below against what you have; if it matches, stop using it and follow the instructions in the firm's or FDA's official notice, or contact the company. If you have health concerns, contact your healthcare provider.
Product codes & lot numbers
UDI-DI: 20199150023312, Lot Numbers: 0232657469 0232670800 0232744945 0232798410 0232862769 0232985251 0233116924 0233116931; UDI-DI: 00199150023318, Lot Numbers: 0232744945
Distribution
Worldwide - US Nationwide distribution.
What does Class II mean?
Class II means use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or the probability of serious harm is remote.
Source: FDA (medical devices), recall Z-2809-2026. Official notice. Republished from public data; RecallFlag is not affiliated with the FDA.