FDA Medical devices recall ·

EOPA Elongated One-Piece Arterial Cannula BLUNT 22FR… recall

Recall number
Z-2807-2026
Date
(initiated June 24, 2026)
Company / brand
Medtronic Perfusion Systems
7611 Northland Dr N, Brooklyn Park, MN, United States
Category
Medical devices › Class II
Classification
Class II
Status
Ongoing
Units affected
11,720 units
Hazard tags
Laceration
Initiated by
Voluntary: Firm initiated

Read the official FDA notice ↗

Product

EOPA Elongated One-Piece Arterial Cannula BLUNT 22FR, intended for use with cardiopulmonary bypass as an arterial return cannula, Model Number 77522

Why was it recalled?

Pinhole leaks may occur in the wire-wound body during cardiopulmonary bypass. Potential risks associated with this condition may include hypovolemia, vessel laceration or perforation, hypotension, ischemia, and organ dysfunction.

What should you do?

FDA enforcement reports do not list a consumer remedy. Check the product codes below against what you have; if it matches, stop using it and follow the instructions in the firm's or FDA's official notice, or contact the company. If you have health concerns, contact your healthcare provider.

Product codes & lot numbers

UDI-DI: 00199150023288, Lot Numbers: 0233286436; UDI-DI: 20199150023282, Lot Numbers: 0232709428 0232744924 0232745061 0232745092 0232798556 0232862763 0232862778 0232862794 0232862798 0232903216 0232903291 0232903383 0232990983 0233286318 0233286414 0233286436 0233427120 0233528730 0233553623 0233556386 0233585113 0233585114 0232670802

Distribution

Worldwide - US Nationwide distribution.

What does Class II mean?

Class II means use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or the probability of serious harm is remote.

Source: FDA (medical devices), recall Z-2807-2026. Official notice. Republished from public data; RecallFlag is not affiliated with the FDA.