FDA Medical devices recall ·

TCD Electrodes (Description recall

Recall number
Z-0368-2027
Date
(initiated October 5, 2026)
Company / brand
Boston Scientific Neuromodulation Corporation
25155 Rye Canyon Loop, Valencia, CA, United States
Category
Medical devices › Class II
Classification
Class II
Status
Ongoing
Units affected
13320
Hazard tags
Foreign material Sterility Microbial contamination Laceration
Initiated by
Voluntary: Firm initiated

Read the official FDA notice ↗

Product

TCD Electrodes (Description/REF(Inner/Outer)): TCD Electrode, Multi-Pack, 3M Cable, 10cm x 28ga (0.36mm)/(TCD-10-3M/TCD-10-3M-P); TCD Electrode, Multi-Pack, 10cm x 28ga (0.36mm)/(TCD-10/TCD-10-P); TCD Electrode, Multi-Pack, 3M Cable, 15cm x 28ga (0.36mm)/(TCD-15-3M/TCD-15-3M-P); TCD Electrode, Multi-Pack, 15cm x 28ga (0.36mm)/(TCD-15/TCD-15-P); TCD Electrode, Multi-Pack, 5cm x 28ga (0.36mm)/(TCD-5/TCD-5-P);

Why was it recalled?

Radiofrequency probes may have small cut marks on the Tyvek side of the sterile pouch, and if the sterile barrier is compromised and devices are used, this could cause microbial or other foreign material contamination to be introduced into patient tissue, which may lead to infection, pain, inflammation, allergic reaction, toxicity, the need for additional medical and/or surgical intervention.

What should you do?

FDA enforcement reports do not list a consumer remedy. Check the product codes below against what you have; if it matches, stop using it and follow the instructions in the firm's or FDA's official notice, or contact the company. If you have health concerns, contact your healthcare provider.

Product codes & lot numbers

Inner REF/UDI-DI/Lot(Expiration): TCD-10-3M/00813250014396/39749333(7/6/2030), 85009822(7/12/2030); TCD-10/00813250010978/39741131(7/5/2030), 39750861(7/6/2030), 39757187(7/7/2030), 39766608(7/8/2030), 39768774(7/8/2030), 39772208(7/9/2030), 39778427(7/9/2030), 39780344(7/9/2030), 39781367(7/10/2030), 85006486(7/12/2030), 85011925(7/13/2030); TCD-15-3M/00813250014402/39770714(7/8/2030); TCD-15/00813250010985/39741129(7/5/2030), 39746187(7/6/2030), 39780349(7/10/2030), 39783782(7/10/2030); TCD-5/00813250010961/39751579(7/7/2030), 39783599(7/10/2030)

Distribution

US Nationwide distribution including in the states of OH, ID, KY, NJ, GA, NY, MO, SC, MN, IA, MI, MA, WI, TX, ME, IN, AR, CO, IL, VT, VA, MS, WA, OR, NV, AZ, UT, FL, WY, CA, OK, NC, TN, MD, DE, LA, HI, CT, WV, AL, NH, ND.

States: Alabama Arkansas Arizona California Colorado Connecticut Delaware Florida Georgia Hawaii Iowa Idaho Illinois Indiana Kentucky Louisiana Massachusetts Maryland Maine Michigan Minnesota Missouri Mississippi North Carolina North Dakota New Hampshire New Jersey Nevada New York Ohio Oklahoma Oregon South Carolina Tennessee Texas Utah Virginia Vermont Washington Wisconsin West Virginia Wyoming

What does Class II mean?

Class II means use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or the probability of serious harm is remote.

Source: FDA (medical devices), recall Z-0368-2027. Official notice. Republished from public data; RecallFlag is not affiliated with the FDA.