FDA Medical devices recall ·

Octopus 4 Tissue Stabilizer, Model 29400 recall

Recall number
Z-2577-2026
Date
(initiated May 12, 2026)
Company / brand
Medtronic Perfusion Systems
7611 Northland Dr N, Brooklyn Park, MN, United States
Category
Medical devices › Class II
Classification
Class II
Status
Ongoing
Units affected
590 units
Initiated by
Voluntary: Firm initiated

Read the official FDA notice ↗

Product

Octopus 4 Tissue Stabilizer, Model 29400

Why was it recalled?

During the manufacturing process, Medtronic personnel identified an assembly issue where the tubing was incorrectly positioned. The short tube was attached to the bottom of the canister, and the long tube was attached to the top, contrary to the intended design.

What should you do?

FDA enforcement reports do not list a consumer remedy. Check the product codes below against what you have; if it matches, stop using it and follow the instructions in the firm's or FDA's official notice, or contact the company. If you have health concerns, contact your healthcare provider.

Product codes & lot numbers

UDI-DI: 00763000543679; Serial Numbers: 0231266463 0231468923 0231478249 0231478253 0231478260 0231651627 0231651634 0231651636 0231651637 0231651638 0231651639 0231651641 0231651642

Distribution

Worldwide - US Nationwide distribution.

What does Class II mean?

Class II means use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or the probability of serious harm is remote.

Source: FDA (medical devices), recall Z-2577-2026. Official notice. Republished from public data; RecallFlag is not affiliated with the FDA.