FDA Medical devices recall ·
EOPA 3D Arterial Cannula DIL TIP 22FR, intended for use with… recall
- Recall number
- Z-2810-2026
- Date
- (initiated June 24, 2026)
- Company / brand
- Medtronic Perfusion Systems
7611 Northland Dr N, Brooklyn Park, MN, United States - Category
- Medical devices › Class II
- Classification
- Class II
- Status
- Ongoing
- Units affected
- 3,621 units
- Hazard tags
- Laceration
- Initiated by
- Voluntary: Firm initiated
Read the official FDA notice ↗
Product
EOPA 3D Arterial Cannula DIL TIP 22FR, intended for use with cardiopulmonary bypass as an arterial return cannula, Model Number 78222
Why was it recalled?
Pinhole leaks may occur in the wire-wound body during cardiopulmonary bypass. Potential risks associated with this condition may include hypovolemia, vessel laceration or perforation, hypotension, ischemia, and organ dysfunction.
What should you do?
FDA enforcement reports do not list a consumer remedy. Check the product codes below against what you have; if it matches, stop using it and follow the instructions in the firm's or FDA's official notice, or contact the company. If you have health concerns, contact your healthcare provider.
Product codes & lot numbers
UDI-DI: 20763000135700, Lot Numbers: 0232602300 0232602383 0232612916 0232680078 0232708967 0232709001 0232910880 0232910881 0232910882 0233077983 0233245850 0233245860 0233245980; UDI-DI: 00763000135706, Lot Numbers: 0232612916 0232709001
Distribution
Worldwide - US Nationwide distribution.
What does Class II mean?
Class II means use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or the probability of serious harm is remote.
Source: FDA (medical devices), recall Z-2810-2026. Official notice. Republished from public data; RecallFlag is not affiliated with the FDA.