FDA Medical devices recall ·

Octopus Evolution AS Tissue Stabilizer, Model TS2500 recall

Recall number
Z-2576-2026
Date
(initiated May 12, 2026)
Company / brand
Medtronic Perfusion Systems
7611 Northland Dr N, Brooklyn Park, MN, United States
Category
Medical devices › Class II
Classification
Class II
Status
Ongoing
Units affected
288 units
Initiated by
Voluntary: Firm initiated

Read the official FDA notice ↗

Product

Octopus Evolution AS Tissue Stabilizer, Model TS2500

Why was it recalled?

During the manufacturing process, Medtronic personnel identified an assembly issue where the tubing was incorrectly positioned. The short tube was attached to the bottom of the canister, and the long tube was attached to the top, contrary to the intended design.

What should you do?

FDA enforcement reports do not list a consumer remedy. Check the product codes below against what you have; if it matches, stop using it and follow the instructions in the firm's or FDA's official notice, or contact the company. If you have health concerns, contact your healthcare provider.

Product codes & lot numbers

UDI-DI: 00643169668904, 00613994764379; Serial Numbers: 0232670681 0232709164 0232709168 0232709194 0232709194 0232709199

Distribution

Worldwide - US Nationwide distribution.

What does Class II mean?

Class II means use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or the probability of serious harm is remote.

Source: FDA (medical devices), recall Z-2576-2026. Official notice. Republished from public data; RecallFlag is not affiliated with the FDA.