FDA Medical devices recall ·

VersaCross Transseptal Sheath Kit (VXS) Product Description/IDs VXS0201… recall

Recall number
Z-2105-2025
Date
(initiated June 10, 2025)
Company / brand
Boston Scientific Corporation
4100 Hamline Ave N, Saint Paul, MN, United States
Category
Medical devices › Class II
Classification
Class II
Status
Ongoing
Units affected
14,724
Hazard tags
Sterility
Initiated by
Voluntary: Firm initiated

Read the official FDA notice ↗

Product

VersaCross Transseptal Sheath Kit (VXS) Product Description/IDs VXS0201 VXS 63CM/45DEG/D0 SHEATH VXS2112 VERSACROSS SHEATH (63CM 45DEG D0)-J Used for the percutaneous introduction of various types of cardiovascular catheters and guidewires to all heart chambers, including the left atrium via transseptal perforation / puncture.

Why was it recalled?

Potential for hole in the Tyvek layer of the sterile barrier pouch, which may compromise device sterility. Issue could potentially lead to systemic infection and sepsis in worst case scenario. Potential punctures may go undetected in clinical settings due to small size and variable location.

What should you do?

FDA enforcement reports do not list a consumer remedy. Check the product codes below against what you have; if it matches, stop using it and follow the instructions in the firm's or FDA's official notice, or contact the company. If you have health concerns, contact your healthcare provider.

Product codes & lot numbers

VXS0201 UDI-DI 00685447020301 Lot 35800622; VXS2112 UDI-DI 00685447019251 Lot 35892045

Distribution

Worldwide distribution - US Nationwide and the countries of Canada, Hong Kong, and Japan.

What does Class II mean?

Class II means use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or the probability of serious harm is remote.

Source: FDA (medical devices), recall Z-2105-2025. Official notice. Republished from public data; RecallFlag is not affiliated with the FDA.