Product
Torflex Transseptal Guiding Sheath Kit (TFK-M) TF85-32-63-45 TORFLEX 8.5F 63CM 45DEG TF85-32-63-55 TORFLEX 8.5F 63CM 55DEG Used for the percutaneous introduction of various types of cardiovascular catheters and guidewires to all heart chambers, including the left atrium via transseptal perforation / puncture.
Why was it recalled?
Potential for hole in the Tyvek layer of the sterile barrier pouch, which may compromise device sterility. Issue could potentially lead to systemic infection and sepsis in worst case scenario. Potential punctures may go undetected in clinical settings due to small size and variable location.
What should you do?
FDA enforcement reports do not list a consumer remedy. Check the product codes below against what you have; if it matches, stop using it and follow the instructions in the firm's or FDA's official notice, or contact the company. If you have health concerns, contact your healthcare provider.
Product codes & lot numbers
TF85-32-63-45 UDI-DI 00685447005391 Lots 35508747 35644132 35746496 35913449; TF85-32-63-55 UDI-DI 00685447005407 Lots 35508748 35738145
Distribution
Worldwide distribution - US Nationwide and the countries of Canada, Hong Kong, and Japan.
What does Class II mean?
Class II means use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or the probability of serious harm is remote.
Source: FDA (medical devices), recall Z-2106-2025. Official notice. Republished from public data; RecallFlag is not affiliated with the FDA.