FDA Medical devices recall ·
BOBBIN VENTILATION TUBE 5PK 1.14 FLPL recall
- Recall number
- Z-0388-2027
- Date
- (initiated September 24, 2026)
- Company / brand
- Medtronic Xomed, Inc.
6743 Southpoint Dr N, Jacksonville, FL, United States - Category
- Medical devices › Class II
- Classification
- Class II
- Status
- Ongoing
- Units affected
- 2000
- Hazard tags
- Sterility
- Initiated by
- Voluntary: Firm initiated
Read the official FDA notice ↗
Product
BOBBIN VENTILATION TUBE 5PK 1.14 FLPL, REF: 1010201, Lot# 0231048277 (GTIN 20763000281612).
Why was it recalled?
Ventilation tubes, that provide a passageway for movement of air between the auditory canal and the middle ear, were not subjected to the required sterilization process, which after implantation could cause infection and subsequent inflammation potentially leading to temporary hearing impairment, otorrhea (fluid discharge) and the need for antibiotics.
What should you do?
FDA enforcement reports do not list a consumer remedy. Check the product codes below against what you have; if it matches, stop using it and follow the instructions in the firm's or FDA's official notice, or contact the company. If you have health concerns, contact your healthcare provider.
Product codes & lot numbers
UDI-DI: 20763000281612. Lots: 0231048277
Distribution
Worldwide - US Nationwide distribution including in the states of NY, AL, VA, MI, AZ, WA, TX, MA, IL, ME, NM, CA, IN, NC, OH, MO, MN, NH, IA, NV, GA, PA, FL, TN, WV, SC, OK, NE, OR, SD, AK, LA, CO, WI, MD, KS, UT, ID, WY, NJ and the countries of Canada, Australia, New Zealand, Hong Kong.
States: Alaska Alabama Arizona California Colorado Florida Georgia Iowa Idaho Illinois Indiana Kansas Louisiana Massachusetts Maryland Maine Michigan Minnesota Missouri North Carolina Nebraska New Hampshire New Jersey New Mexico Nevada New York Ohio Oklahoma Oregon Pennsylvania South Carolina South Dakota Tennessee Texas Utah Virginia Washington Wisconsin West Virginia Wyoming
What does Class II mean?
Class II means use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or the probability of serious harm is remote.
Source: FDA (medical devices), recall Z-0388-2027. Official notice. Republished from public data; RecallFlag is not affiliated with the FDA.