FDA Medical devices recall ·

Boston Scientific Corporation StoneSmart Connect Console Contents: (1)… recall

Recall number
Z-3188-2026
Date
(initiated July 22, 2026)
Company / brand
Boston Scientific Corporation
100 Boston Scientific Way, Marlborough, MA, United States
Category
Medical devices › Class II
Classification
Class II
Status
Ongoing
Units affected
982 units
Initiated by
Voluntary: Firm initiated

Read the official FDA notice ↗

Product

Boston Scientific Corporation StoneSmart Connect Console Contents: (1) Power Cord, (1) Power Adaptor, (1)DVI Cable, (1) DVI to HDMI Cable, (1) DisplayPort Cable REF: M0067933000, M006793300H0, M006793300R0, M006793300D0, M006793300Z0, M0067930000, M006793000R0

Why was it recalled?

A subset of power adapters used with the connect console may exhibit higher-than-expected resistance in the electrical grounding path. As a result, the power adaptor may fail electrical safety testing during installation.

What should you do?

FDA enforcement reports do not list a consumer remedy. Check the product codes below against what you have; if it matches, stop using it and follow the instructions in the firm's or FDA's official notice, or contact the company. If you have health concerns, contact your healthcare provider.

Product codes & lot numbers

All Serial Numbers: REF: M0067933000 (UDI 00191506002510, UDI (00191506021771); REF: M006793300H0 (UDI 00191506002558); REF: M006793300R0 (UDI 00191506002534), UDI (00191506021788); REF: M006793300D0 (UDI 00191506002527); REF: M006793300Z0 (UDI 00191506002541); UPN M0067933100 (UDI 00191506022549)

Distribution

Worldwide - US Nationwide distribution including in the states of AZ, FL, CA, OR, IL, NC, NE, UT, CO, OH, AR, NM, AL, WI, TN, IN, GA, OH, WI, KY, CT, KS, SC, IL, NE, MO, MS, DE, TX, NY, WV, MA, MT, WA, WY, NV, HI, ID, MO, ME, WV, NE, OK, MN, LA, WV, NV, PA, NH, NJ, MI, MO, AK, VA, MS, WY, ME and the countries of Australia, Finland, France, Germany, Great Britain, Greece, Ireland, Italy, Lebanon, Norway, Poland, Portugal, Saudi Arabia, Slovakia, Spain, Sweden, Switzerland, Turkiye, Canada, Hong Kong, Singapore, Taiwan, Argentina, Chile, Japan.

States: Alaska Alabama Arkansas Arizona California Colorado Connecticut Delaware Florida Georgia Hawaii Idaho Illinois Indiana Kansas Kentucky Louisiana Massachusetts Maine Michigan Minnesota Missouri Mississippi Montana North Carolina Nebraska New Hampshire New Jersey New Mexico Nevada New York Ohio Oklahoma Oregon Pennsylvania South Carolina Tennessee Texas Utah Virginia Washington Wisconsin West Virginia Wyoming

What does Class II mean?

Class II means use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or the probability of serious harm is remote.

Source: FDA (medical devices), recall Z-3188-2026. Official notice. Republished from public data; RecallFlag is not affiliated with the FDA.