FDA Medical devices recall ·
NAMIC Angiographic Syringe, labeled as follows: 1… recall
This is an FDA Class I recall — the most serious type.
- Recall number
- Z-1717-2026
- Date
- (initiated February 27, 2026)
- Company / brand
- Medline Industries, LP
3 Lakes Dr, Northfield, IL, United States - Category
- Medical devices › Class I
- Classification
- Class I
- Status
- Ongoing
- Units affected
- 149469 kits
- Initiated by
- Voluntary: Firm initiated
Read the official FDA notice ↗
Product
NAMIC Angiographic Syringe, labeled as follows: 1. SYRINGE,20ML,CONTROL,-,PKG, Medline SKU 70075027; 2. SYRINGE,10ML,TR/FR,RA,W/RES, Medline SKU 70085007; 3. MTO,SYRINGE,10ML,TR/FG,RA,-,PKG, Medline SKU 70085107; 4. SYRINGE,12ML,TR/FR,RA,W/RES, Medline SKU 70087007; 5. MTO,SYRINGE,12ML,TR/FG,RA,W/RES, Medline SKU 70087107; 6. SYRINGE,12ML,TR/FR,RA,W/O,RES,-,PKG, Medline SKU 70088007; 7. SYRINGE,10ML,PP/FR,RA,-,PKG, Medline SKU 70095007; 8. SYRINGE,10ML,PP/FG,RA,-,PKG, Medline SKU 70095107; 9. SYRINGE,12ML,W/RES,PP/FR,RA,PG, Medline SKU 70097007; 10. SYRINGE,12ML,PP/FG,RA,W/RES,-,PKG, Medline SKU 70097107; 11. SYRINGE,12ML,PP/FG,RA,W/O,RES, Medline SKU 70098107; 12. SYRINGE,20ML,TR/FR,RA,OEM,PG, Medline SKU 80075027; 13. SYRINGE,10ML,ANGIO,W/RA,-,OEM, Medline SKU 80085007; 14. SYRINGE,12ML,TR/FR,RA,OEM, Medline SKU 80087007; 15. SYRINGE,10ML,PP/FR,RA,-,OEM,-,PG, Medline SKU 80095007; 16. SYRINGE,10ML,PP/FG,RA,OEM,PG, Medline SKU 80095107; 17. SYRINGE,12ML,PP/FG,RA,-,OEM,-,PG, Medline SKU 80097107; 18. CATH SET UP PACK, Medline SKU DYNJ46711O
Why was it recalled?
Medline Industries, LP has identified through post-market surveillance a potential risk of the syringe rotating adaptor unwinding during use, which may result in a loose connection and/or full disconnection between the syringe and manifold.
What should you do?
FDA enforcement reports do not list a consumer remedy. Check the product codes below against what you have; if it matches, stop using it and follow the instructions in the firm's or FDA's official notice, or contact the company. If you have health concerns, contact your healthcare provider.
Product codes & lot numbers
Medline SKU 60010027, Lot Number 25EBF689; Medline SKU 60010027, Lot Number 25CBL960; Medline SKU 60010027, Lot Number 24LBO743; Medline SKU 60010195, Lot Number 25ABI082; Medline SKU 60010195, Lot Number 24IBJ827; Medline SKU 60010195, Lot Number 24HBO437; Medline SKU 60010195, Lot Number 24DBR149; Medline SKU 60010195, Lot Number 24CBF820; Medline SKU 60011433, Lot Number 26BBJ156; Medline SKU 60011433, Lot Number 26ABH642; Medline SKU 60011433, Lot Number 25KBJ356; Medline SKU 60011433, Lot Number 25HBQ865; Medline SKU 60011433, Lot Number 25FBS131; Medline SKU 60011433, Lot Number 25DBR246; Medline SKU 60011433, Lot Number 25CBS392; Medline SKU 60011433, Lot Number 25CBB739; Medline SKU 60011433, Lot Number 24JBR628; Medline SKU 60011433, Lot Number 24EBL302; Medline SKU 60011433, Lot Number 24BBS083; Medline SKU 60011433, Lot Number 23LBH316; Medline SKU 60011433, Lot Number 23JBP386; Medline SKU 60011433, Lot Number 23HBB765; Medline SKU 60011433, Lot Number 23FBC617; Medline SKU 60011433, Lot Number 23EBD152; Medline SKU 60011433, Lot Number 23DBR429; Medline SKU 60020323, Lot Number 25LBK008; Medline SKU 60020323, Lot Number 25JBJ120; Medline SKU 60020323, Lot Number 25HBQ807; Medline SKU 60020323, Lot Number 25GBA069; Medline SKU 60020323, Lot Number 25EBC643; Medline SKU 60020323, Lot Number 25CBE212; Medline SKU 60023904, Lot Number 26BBB137; Medline SKU 60023904, Lot Number 26ABL323; Medline SKU 60023904, Lot Number 25JBM214; Medline SKU 60023904, Lot Number 25EBM
Distribution
Worldwide distribution: US (nationwide) including Puerto Rico (PR) and OUS (foreign) to countries of: Canada (CA), Netherlands (NL), Australia (AU), Korea, Republic of (KR), Sri Lanka (LK), Pakistan (PK), Japan (JP), United Arab Emirates (AE), Singapore (SG) and Slovakia (SK).
States: California Puerto Rico
What does Class I mean?
Class I is the most serious FDA recall class: there is a reasonable probability that using or being exposed to the product will cause serious adverse health consequences or death.
Source: FDA (medical devices), recall Z-1717-2026. Official notice. Republished from public data; RecallFlag is not affiliated with the FDA.