FDA Medical devices recall ·

Medline convenience kits containing BD Urinary Catheter Trays that include… recall

Recall number
Z-0420-2027
Date
(initiated September 24, 2026)
Company / brand
Medline Industries, LP
3 Lakes Dr, Northfield, IL, United States
Category
Medical devices › Class II
Classification
Class II
Status
Ongoing
Units affected
7,043 kits
Hazard tags
Microbial contamination
Initiated by
Voluntary: Firm initiated

Read the official FDA notice ↗

Product

Medline convenience kits containing BD Urinary Catheter Trays that include CliniClean Pericare Wipes ABDOMINAL HYSTERECTOMY DYNJT8877 LAPAROSCOPIC HYSTERECTOMY DYNJ900244J LAVH DYNJ903392L PELVIC/LAVH/TUBALIGATION T DYNJ910662B WPH MAJOR VAGINAL PACK DYNJ89783 DAVINCI HYSTERECTOMY CDS CDS985344F GYN ABDOMINAL CDS CDS982654L GYN LAP HYSTERECTOMY CDS CDS982662Q LAVH DYNJ909742B

Why was it recalled?

Convenience kits contain additional product codes and lots of recalled CliniClean Pericare Wipes identified by manufacturer. Wipes may be contaminated with a lubricant used in the manufacturing process. Use of affected wipes may cause localized skin effects such as irritation, erythema, or mild inflammation. Failure to perform pericare may increase the risk of localized infection or catheter-associated urinary tract infection.

What should you do?

FDA enforcement reports do not list a consumer remedy. Check the product codes below against what you have; if it matches, stop using it and follow the instructions in the firm's or FDA's official notice, or contact the company. If you have health concerns, contact your healthcare provider.

Product codes & lot numbers

DYNJT8877 UDI-DI 10198459764486 Lots 26EBB204 26GBP480 DYNJ900244J UDI-DI 10198459478727 Lots 26ABB104 26CBN466 DYNJ903392L UDI-DI 10198459522758 Lots 25KBE512 25LBA207 DYNJ910662B UDI-DI 10198459477430 Lot 25LBF442 DYNJ89783 UDI-DI 10198459289323 Lots 25LBB743 26ABN735 CDS985344F UDI-DI 10195327312565 Lots 25LBG617 26ABH355 26ABJ895 CDS982654L UDI-DI 10195327306830 Lot 26EBN087 CDS982662Q UDI-DI 10195327306793 Lots 25LBF560 26ABD832 26ABP323 DYNJ909742B UDI-DI 10198459107221 Lots 25LBD208 26ABO287

Distribution

US Nationwide distribution in the states of AL, AR, CA, FL, GA, IL, IN, LA, MA, MN, MO, MS, MT, NC, NE, NV, NY, OH, OK, RI, SC, TN, TX, WI.

States: Alabama Arkansas California Florida Georgia Illinois Indiana Louisiana Massachusetts Minnesota Missouri Mississippi Montana North Carolina Nebraska Nevada New York Ohio Oklahoma Rhode Island South Carolina Tennessee Texas Wisconsin

What does Class II mean?

Class II means use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or the probability of serious harm is remote.

Source: FDA (medical devices), recall Z-0420-2027. Official notice. Republished from public data; RecallFlag is not affiliated with the FDA.