FDA Medical devices recall ·

Medline medical convenience kits, containing NAMIC Angiographic Rotating… recall

This is an FDA Class I recall — the most serious type.

Recall number
Z-1722-2026
Date
(initiated February 27, 2026)
Company / brand
Medline Industries, LP
3 Lakes Dr, Northfield, IL, United States
Category
Medical devices › Class I
Classification
Class I
Status
Ongoing
Units affected
20 kits
Initiated by
Voluntary: Firm initiated

Read the official FDA notice ↗

Product

Medline medical convenience kits, containing NAMIC Angiographic Rotating Adaptor (RA) Control Syringes (Namic RA syringes), labeled as follows: 1. LAPAROSCOPIC HERNIORRHAPHY DSC, Medline SKU # DYNJ63021B; 2. LAP PACK, Medline SKU # DYNJ64297B; 3. MINOR PROCEDURE PACK, Medline SKU # DYNJ82411.

Why was it recalled?

Medline Industries, LP has identified through post-market surveillance a potential risk of the syringe rotating adaptor unwinding during use, which may result in a loose connection and/or full disconnection between the syringe and manifold.

What should you do?

FDA enforcement reports do not list a consumer remedy. Check the product codes below against what you have; if it matches, stop using it and follow the instructions in the firm's or FDA's official notice, or contact the company. If you have health concerns, contact your healthcare provider.

Product codes & lot numbers

Medline SKU # DYNJ63021B, UDI/DI each 10193489833775UDI/DI case 40193489833776, Lot Number: 26AMD887; Medline SKU # DYNJ64297B, UDI/DI each 10198459472992UDI/DI case 40198459472993, Lot Number: 26AMA970; Medline SKU # DYNJ82411, UDI/DI each 10195327193232UDI/DI case 40195327193233, Lot Number: 25IMB049.

Distribution

Worldwide distribution: US (nationwide) including Puerto Rico (PR) and OUS (foreign) to countries of: Canada (CA), Netherlands (NL), Australia (AU), Korea, Republic of (KR), Sri Lanka (LK), Pakistan (PK), Japan (JP), United Arab Emirates (AE), Singapore (SG) and Slovakia (SK).

States: California Puerto Rico

What does Class I mean?

Class I is the most serious FDA recall class: there is a reasonable probability that using or being exposed to the product will cause serious adverse health consequences or death.

Source: FDA (medical devices), recall Z-1722-2026. Official notice. Republished from public data; RecallFlag is not affiliated with the FDA.