FDA Medical devices recall ·

Medline medical convenience kits, containing NAMIC Angiographic Rotating… recall

This is an FDA Class I recall — the most serious type.

Recall number
Z-1728-2026
Date
(initiated February 27, 2026)
Company / brand
Medline Industries, LP
3 Lakes Dr, Northfield, IL, United States
Category
Medical devices › Class I
Classification
Class I
Status
Ongoing
Units affected
439 kits
Initiated by
Voluntary: Firm initiated

Read the official FDA notice ↗

Product

Medline medical convenience kits, containing NAMIC Angiographic Rotating Adaptor (RA) Control Syringes (Namic RA syringes), labeled as follows: 1. ROBOTIC PACK, Medline SKU # DYNJ58475F; 2. URO/GEN ROBOTIC PACK, Medline SKU # DYNJ60803J.

Why was it recalled?

Medline Industries, LP has identified through post-market surveillance a potential risk of the syringe rotating adaptor unwinding during use, which may result in a loose connection and/or full disconnection between the syringe and manifold.

What should you do?

FDA enforcement reports do not list a consumer remedy. Check the product codes below against what you have; if it matches, stop using it and follow the instructions in the firm's or FDA's official notice, or contact the company. If you have health concerns, contact your healthcare provider.

Product codes & lot numbers

Medline SKU # DYNJ58475F, UDI/DI each 10198459572388, UDI/DI case 40198459572389, Lot Number: 25LMK173; Medline SKU # DYNJ60803J, UDI/DI each 10195327138820, UDI/DI case 40195327138821, Lot Number: 26BMA268; Medline SKU # DYNJ60803J, UDI/DI each 10195327138820, UDI/DI case 40195327138821, Lot Number: 26AMH928; Medline SKU # DYNJ60803J, UDI/DI each 10195327138820, UDI/DI case 40195327138821, Lot Number: 25EMD876; Medline SKU # DYNJ60803J, UDI/DI each 10195327138820, UDI/DI case 40195327138821, Lot Number: 25EMA519; Medline SKU # DYNJ60803J, UDI/DI each 10195327138820, UDI/DI case 40195327138821, Lot Number: 25DMA846; Medline SKU # DYNJ60803J, UDI/DI each 10195327138820, UDI/DI case 40195327138821, Lot Number: 25BMH004; Medline SKU # DYNJ60803J, UDI/DI each 10195327138820, UDI/DI case 40195327138821, Lot Number: 25BMA218; Medline SKU # DYNJ60803J, UDI/DI each 10195327138820, UDI/DI case 40195327138821, Lot Number: 24KMI208; Medline SKU # DYNJ60803J, UDI/DI each 10195327138820, UDI/DI case 40195327138821, Lot Number: 24KMC007; Medline SKU # DYNJ60803J, UDI/DI each 10195327138820, UDI/DI case 40195327138821, Lot Number: 24JMD850; Medline SKU # DYNJ60803J, UDI/DI each 10195327138820, UDI/DI case 40195327138821, Lot Number: 24IMI939; Medline SKU # DYNJ60803J, UDI/DI each 10195327138820, UDI/DI case 40195327138821, Lot Number: 24EDB694; Medline SKU # DYNJ60803J, UDI/DI each 10195327138820, UDI/DI case 40195327138821, Lot Number: 24DDB793; Medline SKU # DYNJ60803J, UDI/DI each 101953

Distribution

Worldwide distribution: US (nationwide) including Puerto Rico (PR) and OUS (foreign) to countries of: Canada (CA), Netherlands (NL), Australia (AU), Korea, Republic of (KR), Sri Lanka (LK), Pakistan (PK), Japan (JP), United Arab Emirates (AE), Singapore (SG) and Slovakia (SK).

States: California Puerto Rico

What does Class I mean?

Class I is the most serious FDA recall class: there is a reasonable probability that using or being exposed to the product will cause serious adverse health consequences or death.

Source: FDA (medical devices), recall Z-1728-2026. Official notice. Republished from public data; RecallFlag is not affiliated with the FDA.