FDA Medical devices recall ·

Tandem Mobi Insulin Pump, Models: 1004004, 1010750, 1012676, 1012719… recall

This is an FDA Class I recall — the most serious type.

Recall number
Z-0316-2027
Date
(initiated September 23, 2026)
Company / brand
Tandem Diabetes Care, Inc.
12400 High Bluff Dr, San Diego, CA, United States
Category
Medical devices › Class I
Classification
Class I
Status
Ongoing
Units affected
80,938 (77,559 US, 3,379 OUS)
Hazard tags
Fall
Initiated by
Voluntary: Firm initiated

Read the official FDA notice ↗

Product

Tandem Mobi Insulin Pump, Models: 1004004, 1010750, 1012676, 1012719, 1013655, 1013656, 1013501, 1013700, 1014081, 1014801, 1014811, 1014813, 1014936, 1016911, 1016912, 1016913, 1016914, 1018075, 1018077, 1018078

Why was it recalled?

Tandem Diabetes Care, Inc. ( Tandem ) has identified that certain Tandem Mobi insulin pumps may generate false-positive occlusion alarms when no actual insulin flow obstruction is present. Investigation determined that increased friction within the drive module can reduce available motor power and, under certain conditions, cause motor speed to fall below the occlusion detection threshold. As a result, the pump may incorrectly detect an occlusion condition and suspend insulin delivery. When an occlusion alarm occurs, insulin delivery remains suspended until the user acknowledges the alarm and resumes therapy. Repeated false-positive occlusion alarms, if not addressed promptly, may result in interruptions of insulin delivery that can increase the risk of elevated blood glucose levels. In severe cases, hyperglycemia may require medical intervention or hospitalization.

What should you do?

FDA enforcement reports do not list a consumer remedy. Check the product codes below against what you have; if it matches, stop using it and follow the instructions in the firm's or FDA's official notice, or contact the company. If you have health concerns, contact your healthcare provider.

Product codes & lot numbers

UDI-DI: 00389152075013 00389152271910 00389152350110 00389152365510 00389152365619 00389152370019 00389152480114 00389152481111 00389152481319. All affected pump models and serial numbers may be affected. Pump serial numbers with frequent false-positive occlusion alarms to be exchanged: 1198599 1199144 1201048 1201777 1202107 1202156 1204500 1204935 1207112 1209405 1209862 1211585 1218143 1220121 1220201 1220470 1220471 1221343 1221347 1221775 1221783 1222111 1223346 1223487 1223488 1223886 1223906 1223940 1224363 1224978 1225017 1225024 1225158 1225171 1225194 1225204 1225589 1225602 1225682 1226359 1226360 1226365 1226671 1226721 1226748 1226859 1226989 1227008 1227021 1227227 1227375 1227442 1229300 1229807 1229969 1230064 1231839 1233326 1233515 1233517 1233562 1233617 1233665 1233701 1233711 1234069 1234080 1234136 1234188 1234240 1234244 1234521 1234597 1234601 1234703 1234715 1234717 1234718 1234732 1235069 1235124 1235131 1235132 1235138 1235193 1235195 1235204 1235335 1236661 1236815 1236825 1236879 1236945 1237266 1237283 1237337 1237722 1237727 1238064 1238085 1238466 1238517 1238533 1238538 1238742 1238801 1238806 1238820 1238851 1238882 1238899 1238923 1239003 1239011 1239016 1239048 1239054 1239057 1239058 1239094 1239104 1239120 1239489 1239496 1239518 1239527 1239560 1239756 1239762 1239794 1239812 1239814 1239815 1239845 1239877 1239884 1239885 1240021 1241967 1242353 1242365 1242366 1242374 1242450 1242747 1243800 1243835 1243889 1243922 1244195 1244451 1244

Distribution

Worldwide - US Nationwide distribution including in the states of CA, FL, NC, OK, UT, WA, IL, PA, MN, VA, NY, IN, HI, OR, GA, MI, KS, CO, WI, SC, null, TX, NE, AZ, OH, NM, NV, MA, SD, IA, TN, ID, MD, RI, CT, NJ, ME, KY, MT, WV, VT, MS, AR, AL, DC, LA, MO, NH, PR, ND, AK, DE, WY and the country of Canada.

States: Alaska Alabama Arkansas Arizona California Colorado Connecticut District of Columbia Delaware Florida Georgia Hawaii Iowa Idaho Illinois Indiana Kansas Kentucky Louisiana Massachusetts Maryland Maine Michigan Minnesota Missouri Mississippi Montana North Carolina North Dakota Nebraska New Hampshire New Jersey New Mexico Nevada New York Ohio Oklahoma Oregon Pennsylvania Puerto Rico Rhode Island South Carolina South Dakota Tennessee Texas Utah Virginia Vermont Washington Wisconsin West Virginia Wyoming

What does Class I mean?

Class I is the most serious FDA recall class: there is a reasonable probability that using or being exposed to the product will cause serious adverse health consequences or death.

Source: FDA (medical devices), recall Z-0316-2027. Official notice. Republished from public data; RecallFlag is not affiliated with the FDA.