FDA Medical devices recall ·
Medline medical convenience kits, containing NAMIC Angiographic Rotating… recall
This is an FDA Class I recall — the most serious type.
- Recall number
- Z-1715-2026
- Date
- (initiated February 27, 2026)
- Company / brand
- Medline Industries, LP
3 Lakes Dr, Northfield, IL, United States - Category
- Medical devices › Class I
- Classification
- Class I
- Status
- Ongoing
- Units affected
- 1698 kits
- Initiated by
- Voluntary: Firm initiated
Read the official FDA notice ↗
Product
Medline medical convenience kits, containing NAMIC Angiographic Rotating Adaptor (RA) Control Syringes (Namic RA syringes), labeled as follows: 1. CSTM MANIFOLD KIT - TITUS, Medline SKU VASC1061; 2. CSTM MANIFOLD KIT - FRYE, Medline SKU VASC1074; 3. CRYO EP MANIFOLD KIT, Medline SKU VASC1198; 4. MANIFOLD KIT - 500PSI OFF, Medline SKU VASCSHPOFF1; 5. MANIFOLD KIT LOW PRESSURE ON, Medline SKU VASCSLPON1.
Why was it recalled?
Medline Industries, LP has identified through post-market surveillance a potential risk of the syringe rotating adaptor unwinding during use, which may result in a loose connection and/or full disconnection between the syringe and manifold.
What should you do?
FDA enforcement reports do not list a consumer remedy. Check the product codes below against what you have; if it matches, stop using it and follow the instructions in the firm's or FDA's official notice, or contact the company. If you have health concerns, contact your healthcare provider.
Product codes & lot numbers
Medline SKU VASC1061, UDI/DI each 10889942464722, UDI/DI case 40889942464723, Lot Number 25IBN161; Medline SKU VASC1061, UDI/DI each 10889942464722, UDI/DI case 40889942464723, Lot Number 25GBF769; Medline SKU VASC1074, UDI/DI each 10889942611751, UDI/DI case 40889942611752, Lot Number 25HBN798; Medline SKU VASC1074, UDI/DI each 10889942611751, UDI/DI case 40889942611752, Lot Number 25GBL527; Medline SKU VASC1198, UDI/DI each 10193489664263, UDI/DI case 40193489664264, Lot Number 25GDA142; Medline SKU VASCSHPOFF1, UDI/DI each 10889942405503, UDI/DI case 40889942405504, Lot Number 25IBP989; Medline SKU VASCSHPOFF1, UDI/DI each 10889942405503, UDI/DI case 40889942405504, Lot Number 25GBL526; Medline SKU VASCSLPON1, UDI/DI each 10889942405510, UDI/DI case 40889942405511, Lot Number 25GBL070.
Distribution
Worldwide distribution: US (nationwide) including Puerto Rico (PR) and OUS (foreign) to countries of: Canada (CA), Netherlands (NL), Australia (AU), Korea, Republic of (KR), Sri Lanka (LK), Pakistan (PK), Japan (JP), United Arab Emirates (AE), Singapore (SG) and Slovakia (SK).
States: California Puerto Rico
What does Class I mean?
Class I is the most serious FDA recall class: there is a reasonable probability that using or being exposed to the product will cause serious adverse health consequences or death.
Source: FDA (medical devices), recall Z-1715-2026. Official notice. Republished from public data; RecallFlag is not affiliated with the FDA.