VERIFY Biological/Integrator Challenge Pack Model recall
- Recall number
- Z-0070-2027
- Date
- (initiated August 31, 2026)
- Company / brand
- Steris Corporation
5960 Heisley Rd, Mentor, OH, United States - Category
- Medical devices › Class II
- Classification
- Class II
- Status
- Ongoing
- Units affected
- 266
- Hazard tags
- Sterility Fall
- Initiated by
- Voluntary: Firm initiated
Read the official FDA notice ↗
Product
VERIFY Biological/Integrator Challenge Pack Model/Catalog Number: LCB006
Why was it recalled?
Geobacillus stearothermophilus spore population in affected products may fall below minimum shelf life specification. When used to monitor steam sterilization cycles, the potential hazard is that the biological indicator will show no growth, indicating that the steam cycle was effective, even if the steam cycle did not achieve the intended spore reduction
What should you do?
FDA enforcement reports do not list a consumer remedy. Check the product codes below against what you have; if it matches, stop using it and follow the instructions in the firm's or FDA's official notice, or contact the company. If you have health concerns, contact your healthcare provider.
Product codes & lot numbers
UDI: 10724995220324 Lot numbers: 260810 - expiration date 08/10/2026 260810A - expiration date 08/10/2026 261106 - expiration date 11/06/2026 270116 - expiration date 01/16/2027
Distribution
Worldwide - US Nationwide distribution.
What does Class II mean?
Class II means use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or the probability of serious harm is remote.
Source: FDA (medical devices), recall Z-0070-2027. Official notice. Republished from public data; RecallFlag is not affiliated with the FDA.