FDA Medical devices recall ·

RAPICIDE PA Part A High-Level Disinfectant, Model recall

Recall number
Z-0155-2027
Date
(initiated September 21, 2026)
Company / brand
Steris Corporation
5960 Heisley Rd, Mentor, OH, United States
Category
Medical devices › Class II
Classification
Class II
Status
Ongoing
Units affected
11806 units
Hazard tags
Injury risk
Initiated by
Voluntary: Firm initiated

Read the official FDA notice ↗

Product

RAPICIDE PA Part A High-Level Disinfectant, Model/Catalog Number: ML020116

Why was it recalled?

If the vented cap becomes blocked, gas may accumulate and the bottle may swell due to increased internal pressure. In some cases, excess vapor or liquid chemistry may escape from the bottle when the cap is twisted during opening, which could result in user exposure if required personal protective equipment (PPE) is not properly used. Such exposure may result in injury to the user or others in close proximity.

What should you do?

FDA enforcement reports do not list a consumer remedy. Check the product codes below against what you have; if it matches, stop using it and follow the instructions in the firm's or FDA's official notice, or contact the company. If you have health concerns, contact your healthcare provider.

Product codes & lot numbers

UDI: 00724995245955 Lot numbers with expiration dates: 689171 5/4/2027 689981 5/12/2027 690993 5/26/2027 691739 6/2/2027 692570 6/11/2027 694357 7/1/2027

Distribution

Worldwide - US Nationwide distribution.

What does Class II mean?

Class II means use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or the probability of serious harm is remote.

Source: FDA (medical devices), recall Z-0155-2027. Official notice. Republished from public data; RecallFlag is not affiliated with the FDA.