FDA Medical devices recall ·
RAPICIDE PA Part A High-Level Disinfectant, Model recall
- Recall number
- Z-0154-2027
- Date
- (initiated September 21, 2026)
- Company / brand
- Steris Corporation
5960 Heisley Rd, Mentor, OH, United States - Category
- Medical devices › Class II
- Classification
- Class II
- Status
- Ongoing
- Units affected
- 374 units
- Hazard tags
- Injury risk
- Initiated by
- Voluntary: Firm initiated
Read the official FDA notice ↗
Product
RAPICIDE PA Part A High-Level Disinfectant, Model/Catalog Number: ML020115
Why was it recalled?
If the vented cap becomes blocked, gas may accumulate and the bottle may swell due to increased internal pressure. In some cases, excess vapor or liquid chemistry may escape from the bottle when the cap is twisted during opening, which could result in user exposure if required personal protective equipment (PPE) is not properly used. Such exposure may result in injury to the user or others in close proximity.
What should you do?
FDA enforcement reports do not list a consumer remedy. Check the product codes below against what you have; if it matches, stop using it and follow the instructions in the firm's or FDA's official notice, or contact the company. If you have health concerns, contact your healthcare provider.
Product codes & lot numbers
UDI: 00724995245658 Lot numbers with expiration dates: 687046 10/5/2027 688017 10/19/2027
Distribution
Worldwide - US Nationwide distribution.
What does Class II mean?
Class II means use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or the probability of serious harm is remote.
Source: FDA (medical devices), recall Z-0154-2027. Official notice. Republished from public data; RecallFlag is not affiliated with the FDA.