FDA Medical devices recall ·

MEDLINE Medical Procedure Kits labeled as: TONSIL PACK CATH… recall

Recall number
Z-2494-2026
Date
(initiated May 4, 2026)
Company / brand
Medline Industries, LP
3 Lakes Dr, Northfield, IL, United States
Category
Medical devices › Class II
Classification
Class II
Status
Ongoing
Units affected
384 units
Initiated by
Voluntary: Firm initiated

Read the official FDA notice ↗

Product

MEDLINE Medical Procedure Kits labeled as: TONSIL PACK CATH, Medline Kit Number/SKU DYNJ65199A

Why was it recalled?

During an internal review, Medline identified representative samples did not meet applicable performance specifications. Identified issues include irregularities in catheter surface finish, instances where the catheter connector detached prematurely from the catheter funnel, and limited cases where the catheter outer diameter did not meet specification.

What should you do?

FDA enforcement reports do not list a consumer remedy. Check the product codes below against what you have; if it matches, stop using it and follow the instructions in the firm's or FDA's official notice, or contact the company. If you have health concerns, contact your healthcare provider.

Product codes & lot numbers

UDI/DI 10195327119751 (each), 40195327119752 (case), Lot Numbers: 23HMI132, 23HMA956 23GMA525, 23DMF610, 23AMH680, 22CBF155.

Distribution

Worldwide distribution - US Nationwide and the countries of AE, BB, BM, BS, CA, CY, GU, NL, PA, QA, SA, VI.

States: California Pennsylvania

What does Class II mean?

Class II means use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or the probability of serious harm is remote.

Source: FDA (medical devices), recall Z-2494-2026. Official notice. Republished from public data; RecallFlag is not affiliated with the FDA.