FDA Medical devices recall ·
MEDLINE Medical Procedure Kits labeled as: 1) BR L&D PACK… recall
- Recall number
- Z-2501-2026
- Date
- (initiated May 4, 2026)
- Company / brand
- Medline Industries, LP
3 Lakes Dr, Northfield, IL, United States - Category
- Medical devices › Class II
- Classification
- Class II
- Status
- Ongoing
- Units affected
- 24388 units
- Initiated by
- Voluntary: Firm initiated
Read the official FDA notice ↗
Product
MEDLINE Medical Procedure Kits labeled as: 1) BR L&D PACK, Medline Kit Number/SKU DYNJ47625B; 2) GYN VAGINAL, Medline Kit Number/SKU DYNJ912219; 3) KIT MAJ VAG SET UP, Medline Kit Number/SKU DYNJ907033B; 4) MINOR VAG PACK, Medline Kit Number/SKU DYNJ68663B; 5) OB DELIVERY PACK, Medline Kit Number/SKU DYNJ14099F; 6) OB DELIVERY PACK, Medline Kit Number/SKU DYNJ14099G; 7) OB PACK, Medline Kit Number/SKU DYNJ03387B; 8) OB PACK, Medline Kit Number/SKU DYNJ03387C; 9) OB PACK, Medline Kit Number/SKU DYNJ53543B; 10) RR-D/E PACK, Medline Kit Number/SKU DYNJ62645D; 11) RR-D/E PACK, Medline Kit Number/SKU DYNJ62645F; 12) RR-D/E PACK, Medline Kit Number/SKU DYNJ62645G; 13) VAGINAL DELIVERY PACK, Medline Kit Number/SKU DYNJ04454K; 14) VAGINAL DELIVERY PACK, Medline Kit Number/SKU DYNJ04454L; 15) VAGINAL PACK, Medline Kit Number/SKU DYNJ01830G.
Why was it recalled?
During an internal review, Medline identified representative samples did not meet applicable performance specifications. Identified issues include irregularities in catheter surface finish, instances where the catheter connector detached prematurely from the catheter funnel, and limited cases where the catheter outer diameter did not meet specification.
What should you do?
FDA enforcement reports do not list a consumer remedy. Check the product codes below against what you have; if it matches, stop using it and follow the instructions in the firm's or FDA's official notice, or contact the company. If you have health concerns, contact your healthcare provider.
Product codes & lot numbers
Medline Kit Number/SKU DYNJ47625B, UDI/DI 10889942182589 (each), 40889942182580 (case), Lot Number 21LBI584; Medline Kit Number/SKU DYNJ912219, UDI/DI 10198459676048 (each), 40198459676049 (case), Lot Number 26BBQ085; Medline Kit Number/SKU DYNJ907033B, UDI/DI 10195327205980 (each), 40195327205981 (case), Lot Number 23BBK437; Medline Kit Number/SKU DYNJ907033B, UDI/DI 10195327205980 (each), 40195327205981 (case), Lot Number 22LBP927; Medline Kit Number/SKU DYNJ907033B, UDI/DI 10195327205980 (each), 40195327205981 (case), Lot Number 22KBA970; Medline Kit Number/SKU DYNJ907033B, UDI/DI 10195327205980 (each), 40195327205981 (case), Lot Number 22JBH943; Medline Kit Number/SKU DYNJ907033B, UDI/DI 10195327205980 (each), 40195327205981 (case), Lot Number 22HBR879; Medline Kit Number/SKU DYNJ68663B, UDI/DI 10195327408886 (each), 40195327408887 (case), Lot Number 25ABD038; Medline Kit Number/SKU DYNJ68663B, UDI/DI 10195327408886 (each), 40195327408887 (case), Lot Number 24IBP599; Medline Kit Number/SKU DYNJ68663B, UDI/DI 10195327408886 (each), 40195327408887 (case), Lot Number 24HBT938; Medline Kit Number/SKU DYNJ68663B, UDI/DI 10195327408886 (each), 40195327408887 (case), Lot Number 24GBB724; Medline Kit Number/SKU DYNJ68663B, UDI/DI 10195327408886 (each), 40195327408887 (case), Lot Number 24DBK110; Medline Kit Number/SKU DYNJ68663B, UDI/DI 10195327408886 (each), 40195327408887 (case), Lot Number 24ABN322; Medline Kit Number/SKU DYNJ68663B, UDI/DI 10195327408886 (each), 4019532740888
Distribution
Worldwide distribution - US Nationwide and the countries of AE, BB, BM, BS, CA, CY, GU, NL, PA, QA, SA, VI.
States: California Pennsylvania
What does Class II mean?
Class II means use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or the probability of serious harm is remote.
Source: FDA (medical devices), recall Z-2501-2026. Official notice. Republished from public data; RecallFlag is not affiliated with the FDA.