MEDLINE Medical Procedure Kits labeled as: 1) RETINA LAC/USC CDS… recall
- Recall number
- Z-2507-2026
- Date
- (initiated May 4, 2026)
- Company / brand
- Medline Industries, LP
3 Lakes Dr, Northfield, IL, United States - Category
- Medical devices › Class II
- Classification
- Class II
- Status
- Ongoing
- Units affected
- 2340 units
- Initiated by
- Voluntary: Firm initiated
Read the official FDA notice ↗
Product
MEDLINE Medical Procedure Kits labeled as: 1) RETINA LAC/USC CDS, Medline Kit Number/SKU CDS983775I; 2) RETINA LAC/USC CDS, Medline Kit Number/SKU CDS983775J; 3) RETINA LAC/USC CDS, Medline Kit Number/SKU CDS983775K; 4) RETINA LAC/USC CDS, Medline Kit Number/SKU CDS983775L; 5) RETINA LAC/USC CDS, Medline Kit Number/SKU CDS983775M; 6) TONSIL PACK, Medline Kit Number/SKU DYNJ84480; 7) VITRECTOMY, Medline Kit Number/SKU DYNJVB1080; 8) VITRECTOMY, Medline Kit Number/SKU DYNJVB1080A.
Why was it recalled?
During an internal review, Medline identified representative samples did not meet applicable performance specifications. Identified issues include irregularities in catheter surface finish, instances where the catheter connector detached prematurely from the catheter funnel, and limited cases where the catheter outer diameter did not meet specification.
What should you do?
FDA enforcement reports do not list a consumer remedy. Check the product codes below against what you have; if it matches, stop using it and follow the instructions in the firm's or FDA's official notice, or contact the company. If you have health concerns, contact your healthcare provider.
Product codes & lot numbers
Medline Kit Number/SKU CDS983775I: UDI/DI 10193489980660 (each), 40193489980661 (case), Lot Number 24BBF438; Medline Kit Number/SKU CDS983775I: UDI/DI 10193489980660 (each), 40193489980661 (case), Lot Number 23LBR516; Medline Kit Number/SKU CDS983775I: UDI/DI 10193489980660 (each), 40193489980661 (case), Lot Number 23LBI385; Medline Kit Number/SKU CDS983775I: UDI/DI 10193489980660 (each), 40193489980661 (case), Lot Number 23GBW083; Medline Kit Number/SKU CDS983775I: UDI/DI 10193489980660 (each), 40193489980661 (case), Lot Number 23GBT137; Medline Kit Number/SKU CDS983775I: UDI/DI 10193489980660 (each), 40193489980661 (case), Lot Number 23GBA705; Medline Kit Number/SKU CDS983775I: UDI/DI 10193489980660 (each), 40193489980661 (case), Lot Number 23EBT206; Medline Kit Number/SKU CDS983775I: UDI/DI 10193489980660 (each), 40193489980661 (case), Lot Number 23BBN853; Medline Kit Number/SKU CDS983775I: UDI/DI 10193489980660 (each), 40193489980661 (case), Lot Number 23BBG437; Medline Kit Number/SKU CDS983775I: UDI/DI 10193489980660 (each), 40193489980661 (case), Lot Number 23BBG444; Medline Kit Number/SKU CDS983775I: UDI/DI 10193489980660 (each), 40193489980661 (case), Lot Number 23BBA852; Medline Kit Number/SKU CDS983775I: UDI/DI 10193489980660 (each), 40193489980661 (case), Lot Number 23BBA859; Medline Kit Number/SKU CDS983775I: UDI/DI 10193489980660 (each), 40193489980661 (case), Lot Number 22IBR361; Medline Kit Number/SKU CDS983775I: UDI/DI 10193489980660 (each), 40193489980661 (ca
Distribution
Worldwide distribution - US Nationwide and the countries of AE, BB, BM, BS, CA, CY, GU, NL, PA, QA, SA, VI.
States: California Pennsylvania
What does Class II mean?
Class II means use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or the probability of serious harm is remote.