FDA Medical devices recall ·
AMERICAN CONTRACT SYSTEMS, Medical Procedure kits and trays labeled as: 1)… recall
- Recall number
- Z-2466-2025
- Date
- (initiated July 14, 2025)
- Company / brand
- American Contract Systems Inc
85 Shaffer Park Dr, Tiffin, OH, United States - Category
- Medical devices › Class II
- Classification
- Class II
- Status
- Ongoing
- Units affected
- 1127 units
- Hazard tags
- Sterility
- Initiated by
- Voluntary: Firm initiated
Read the official FDA notice ↗
Product
AMERICAN CONTRACT SYSTEMS, Medical Procedure kits and trays labeled as: 1) CYSTO, Model/Item Number ASCY24A; 2) CYSTO PACK, Model/Item Number BUCY78E; 3) CYSTO PACK, Model/Item Number LMCP22T; 4) TUR BASIN PACK, Model/Item Number LMTU31T; 5) CYSTO, Model/Item Number SACY80S; 6) UROLOGY PACK, Model/Item Number SPUR64F.
Why was it recalled?
During an internal investigation, ACS determined that the rationale for sterilization of certain components was not justified and therefore, some components are not fit for further sterilization according to ACS procedures. As a result, ACS is unable to confirm product sterilization assurance requirements were met.
What should you do?
FDA enforcement reports do not list a consumer remedy. Check the product codes below against what you have; if it matches, stop using it and follow the instructions in the firm's or FDA's official notice, or contact the company. If you have health concerns, contact your healthcare provider.
Product codes & lot numbers
1) Model/Item Number ASCY24A: UDI/DI 00191072201621, Kit Lot Numbers: 12-7370111, 12-7388611, 12-7388611, 12-7139611, 12-7191911, 12-7231611, 12-7263911; 2) Model/Item Number BUCY78E: UDI/DI 00191072235350, Kit Lot Numbers: 42-7962811; 3) Model/Item Number LMCP22T: UDI/DI 00191072235510, Kit Lot Numbers: 42-8027411, 42-8106111, 42-7885511, 42-7823511; 4) Model/Item Number LMTU31T: UDI/DI 00191072235527, Kit Lot Numbers: 42-7987311, 42-8059311, 42-8112211, 42-7832311, 42-7845011, 42-7914911; 5) Model/Item Number SACY80S: UDI/DI 00191072233370, Kit Lot Numbers: 70-052954, 70-052248, 70-51139, 70-053309, 70-053805; 6) Model/Item Number SPUR64F: UDI/DI 00191072238764, Kit Lot Numbers: 12-7394711, 12-7328411, 12-7096811, 12-7168211.
Distribution
US distribution to states of: IL and LA
What does Class II mean?
Class II means use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or the probability of serious harm is remote.
Source: FDA (medical devices), recall Z-2466-2025. Official notice. Republished from public data; RecallFlag is not affiliated with the FDA.