FDA Medical devices recall ·
Medical convenience kits containing BD CliniClean Pericare Wipes within BD… recall
- Recall number
- Z-3292-2026
- Date
- (initiated August 10, 2026)
- Company / brand
- American Contract Systems Inc
85 Shaffer Park Dr, Tiffin, OH, United States - Category
- Medical devices › Class II
- Classification
- Class II
- Status
- Ongoing
- Units affected
- 1,804 kits total
- Hazard tags
- Microbial contamination
- Initiated by
- Voluntary: Firm initiated
Read the official FDA notice ↗
Product
Medical convenience kits containing BD CliniClean Pericare Wipes within BD Urinary Catheter Trays ROBOTIC PACK - 242561 Model No. MHRB88AO
Why was it recalled?
Kits contain manufacturer-recalled perineal care (pericare) wipes within urinary catheter trays, which may be contaminated with a lubricant used during the manufacturing process. Use of a contaminated wipe may lead to localized skin effects such as irritation, erythema, or mild inflammation. Failure to perform pericare due to disposal of contaminated wipes may increase risk of localized infection or catheter-associated urinary tract infection (CAUTI).
What should you do?
FDA enforcement reports do not list a consumer remedy. Check the product codes below against what you have; if it matches, stop using it and follow the instructions in the firm's or FDA's official notice, or contact the company. If you have health concerns, contact your healthcare provider.
Product codes & lot numbers
UDI-DI 191072236531 Lots 42-8513411 42-8580011 42-8663211 42-8775711 42-8801311 42-8840611 42-8865911
Distribution
US Nationwide distribution in the states of Iowa, Minnesota, Wisconsin, and South Dakota.
States: Iowa Minnesota South Dakota Wisconsin
What does Class II mean?
Class II means use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or the probability of serious harm is remote.
Source: FDA (medical devices), recall Z-3292-2026. Official notice. Republished from public data; RecallFlag is not affiliated with the FDA.