FDA Medical devices recall ·
REF: MIDLINE CATHETER KIT: 3 FR X 20 CM/CDC-02031-MK1A… recall
This is an FDA Class I recall — the most serious type.
- Recall number
- Z-2394-2026
- Date
- (initiated April 21, 2026)
- Company / brand
- Arrow International, LLC
3015 Carrington Mill Blvd, Morrisville, NC, United States - Category
- Medical devices › Class I
- Classification
- Class I
- Status
- Ongoing
- Units affected
- 177,014
- Hazard tags
- Sterility Injury risk
- Initiated by
- Voluntary: Firm initiated
Read the official FDA notice ↗
Product
Description/REF: MIDLINE CATHETER KIT: 3 FR X 20 CM/CDC-02031-MK1A; AM/AT PI MIDLINE 1L: 4.5FR X 15CM/CDC-41541-MPK1A; AM/AT PI MIDLINE 2L: 5.5FR X 15CM/CDC-41552-MPK1A; AGBA PI MIDLINE 1L: 4.5FR X 15CM/ASK-41541-VFHM; AGBA PI MIDLINE 1-L: 4.5FR X 15CM/CDC-41541-MPK; AGBA PI MIDLINE 1-L: 4.5FR X 15CM/CDC-41541-MPKB; AGBA PI MIDLINE 1-L: 4.5FR X 15CM/CDC-41541-MPKC; AGBA PI MIDLINE 2-L: 5.5FR X 15CM/CDC-41552-MPK; AGBA PI MIDLINE 2-L: 5.5FR X 15CM/CDC-41552-MPKB; AGBA PI MIDLINE 2-L: 5.5FR X 15CM/CDC-41552-MPKC; PI MIDLINE 1-L: 4FR X 20CM/CDC-32041-MPK; PI MIDLINE 1-L: 4FR X 20CM/CDC-32041-MPKB; PI MIDLINE 1L: 4FR X 20CM/CDC-32041-MPKC; PI MIDLINE 2-L: 5FR X 20CM/CDC-32052-MPKB; PI PICC NaviCurve: 1L 4FR x 55CM w BP/DLX-35541-CURVB; PI PICC: 1L 4FR x 55CM w Biopatch/DLX-35541-HPKB; PI PICC: 1L 4FR x 55CM w CHG/DLX-35541-HPKC; PI PICC: 2L 5FR x 55CM w Biopatch/DLX-35552-HPKB; PI PICC: 2L 5FR X 55CM w CHG/DLX-35552-HPKC; PI PICC: 3L 6FR x 55CM w Biopatch/DLX-35563-HPKB; AGBA PICC/DELTA KIT: 1L 4.5 FR X 40 CM/ASK-44041-NS; AGBA PICC/DELTA FG: 2L 5.5 FR X 40 CM/ASK-44052-NS; AGBA PICC/DELTA KIT: 1L 4.5 FR X 50 CM/ASK-45041-NS; PI PICC KIT: 2L 5.5 FR X 50 CM/ASK-45052-TG; PI AGBA PICC KIT: 3L 6 FR X 50 CM W/VPS/ASK-45063-NS; PICC KIT: 1L 4.5 FR X 55 CM CG+ VPS/ASK-45541-RH; AGBA PICC: 1L 4.5FR x 55CM w Teg CHG/DLX-45541-HPKC; AGBA PICC: 2L 5.5FR x 55CM w Teg CHG/DLX-45552-HPKC; AGBA PICC NaviCurve: 3L 6FR x 55CM TCG/DLX-45563-CURVC; AGBA PICC: 3L 6FR x 55CM w Biopatch/DLX-45563-HPKB; AGBA PICC: 3L 6FR x 55CM w Teg CHG/DLX-45563-HPKC; PI PICC KIT: 1-L 4 FR X 55 CM TIPTRACKER/CDC-35541-TTS; PI PICC NaviCurve: 1L 4FR x 55CM w CHG/DLX-35541-CURVC; PI PICC G4 Stylet: 1L 4FR x 55CM w BP/DLX-35541-VPSB; PI PICC NaviCurve: 2L 5FR x 55CM w CHG/DLX-35552-CURVC; PI PICC G4 Stylet: 2L 5FR x 55CM w BP/DLX-35552-VPSB; PI PICC G4 Stylet: 2L 5FR x 55CM w CHG/DLX-35552-VPSC; PI PICC NaviCurve: 3L 6FR x 55CM w CHG/DLX-35563-CURVC; PI PICC G4 Stylet: 3L 6FR x 55CM w CHG/DLX-35563-VPSC; AGBA PICC/DELTA KIT: 1-L 4.5 FR X 40 CM/CDC-44041-VPS2; PICC KIT: 1L 4.5 FR X 55 CM VPS/ASK-45541-UWH1; PI AGBA PICC KIT: 2L 5.5FR X 55CM W/ VPS/ASK-45552-UWH1; AGBA PICC G4 Stylet: 1L 4.5FR x 55CM BP/DLX-45541-VPSB; ACCESS TRAY/ASK-04001-NWM; ACCESS TRAY/ASK-04001-VCUH3; PI PICC:1L 4FR X 40CM PRELOAD VPS STYLET/CDC-34041-VPS; PI PICC:2L 5FR X 50CM PRELOAD VPS STYLET/CDC-35052-VPS; ACCESS KIT/ASK-04001-BWK2; ARROW(R) VPS(R) ACCESS KIT/ASK-04001-DU11; MULTI LUMEN PI CVC KIT: 3L 7 FR X 6 IN/ASK-42703-PVD; MULTI LUMEN PI CVC KIT: 3L 7 FR X 16CM/ASK-12703-LMDTG1; MULTI LUMEN PI CVC KIT: 3L 7 FR X 8IN/ASK-15703-ECMC4; MULTI LUMEN PI CVC KIT: 3L 7 FR X 20 CM/ASK-15703-PNM; CVC KIT: 3 LUMEN 12 FR X 6IN/ASK-22123-UAB; 2-L CVC KIT: 4 FR X13 CM/CDC-24402-X1A; LBCVC KIT: 3L 12 FR X 20 CM AGB/CDC-25123-X1A; CVC KIT: 3L 7 FR X 16 CM/ASK-42703-CPR; CVC KIT: 2L 8 FR X 6 IN (16 CM) AGB+/ASK-42802-DMC2; CVC KIT: 2 LUMEN 8 FR X 8 IN (20 CM)/ASK-45802-PCMH1; PI CVC KIT: 3L 7 FR X 6 IN (16 CM) AGB+/ASK-42703-MM1; PI CVC KIT: 3L 7 FR X 16 CM AGB+/ASK-42703-PHUMC1; PI CVC KIT: 3L 7 FR X 16 CM AGB+/ASK-42703-PMSG1; MULTILUMEN PI CVC KIT: 3L 7 FR X 6/ASK-42703-POSU3; PI CVC KIT: 3L 7 FR X 16 CM AGB+/ASK-42703-PSAU1; PI CVC KIT: 3L 7 FR X 16 CM AGB+/ASK-42703-RHR1; PI CATHETER 2L: 8 FR X 16 CM AGB+/ASK-42802-DU1; PI CVC KIT: 2L 8 FR X 6 IN/ASK-42802-PHF4; PI CVC KIT: 2L 8 FR X 16 CM AGB+/ASK-42802-PHHN1; MULTILUMEN CVC KIT: 2L 8 FR X 6 IN/ASK-42802-POSU3; PI CVC KIT: 2-L 8 FR X 16 CM AGB+/ASK-42802-PUCD1; MULTI-LUMEN CVC KIT:2-L 8FR X 6IN (16CM)/ASK-42802-PWHC1; PI CVC KIT: 2L 8 FR X 16 CM AGB+/ASK-42802-UHC; PI CVC KIT: 4L 8.5 FR X 16 CM AGB+/ASK-42854-MCY1; PI CVC KIT: 3L 7FR X 8IN (20 CM) AGB+/ASK-45703-BCH; PI CVC KIT: 3L 7 FR X 20 CM AGB+/ASK-45703-EMC; PI CVC KIT: 3 LUMEN 7 FR X 8 IN (20 CM)/ASK-45703-MHMC1; PI CVC KIT: 3-L 7 FR X 20 CM AGB+/ASK-45703-NWM; MULTILUMEN PI CVC KIT: 3L 7 FR X 20 CM/ASK-45703-PCOR; MULTILUMEN PI CVC KIT: 3L 7 FR X 20 CM/ASK-45703-PIMH; MULTI LUMEN CVC KIT: 3L
Why was it recalled?
Lidocaine, Bupivacaine, and 0.9% sodium chloride, the subject of a supplier recall, were included into kits and sets. The supplier was found to have quality issues and limited reports of drug ineffectiveness, and if sterility of a water-based injectable product is compromised this could lead to serious or life-threatening injuries.
What should you do?
FDA enforcement reports do not list a consumer remedy. Check the product codes below against what you have; if it matches, stop using it and follow the instructions in the firm's or FDA's official notice, or contact the company. If you have health concerns, contact your healthcare provider.
Product codes & lot numbers
REF(Material)/UDI-DI/Lot(Batch): CDC-02031-MK1A/10801902193275/33F25C0444; CDC-41541-MPK1A/10801902193275/33F24L1137; CDC-41552-MPK1A/10801902193275, 50801902130162/33F24L0382, 33F25A0315, 33F25B0630, 33F25K0421, 33S24B0571; ASK-41541-VFHM/10801902193275, 10801902218862/33F24L1080, 33F25B0475, 33F25C0351, 33F25D0404, 33F25E0556, 33F25E0784, 33F25F0186, 33F25J0218, 33F25K0077, 33F25K0100, 33F25M0182, 33F26A1549, 33F26B0823; CDC-41541-MPK/10801902193275, 10801902193275/33F25H0768, 33S23K0382; CDC-41541-MPKB/10801902193275, 10801902205411/33F24H0429, 33F24K0043, 33F24M0067, 33F24M0149, 33F25A0390, 33F25B0099, 33F25B0343, 33F25C0578, 33F25E0421, 33F25F0401, 33F25G0515, 33F25G0710, 33F25H0817, 33F25J1008, 33F25L0015, 33F25L0050, 33F25L0295, 33F25M0499, 33F26A0330, 33X25J0003; CDC-41541-MPKC/10801902205435, 10801902193275/33F24D0982, 33F24F0166, 33F24F0167, 33F24H0398, 33F24H0454, 33F24J0380, 33F24J0389, 33F24K0581, 33F24L0182, 33F24M0022, 33F24M0077, 33F25A0348, 33F25B0228, 33F25B0281, 33F25B0289, 33F25B0344, 33F25C0513, 33F25C0532, 33F25C0558, 33F25D0710, 33F25D0736, 33F25F0020, 33F25F0025, 33F25F0032, 33F25H0818, 33F25J0577, 33F25J0657, 33F25L0251, 33F25L0296, 33S25A0211, 33F26C0435; CDC-41552-MPK/10801902201505/33F24D0561, 33F26B0490, 33S24A0521; CDC-41552-MPKB/10801902193275, 10801902205428/33F24J0183, 33F24J0221, 33F24L0010, 33F24L0173, 33F24M0114, 33F25B0223, 33F25B0299, 33F25D0724, 33F25F0436, 33F25G0640, 33F25K0354, 33F25L0265, 33F26B1048; CDC-41552-MPKC/10801902205442, 10
Distribution
US Nationwide distribution including in the states of VA, NY, UT, SC, LA, MN, PA, AZ, CA, MI, OH, FL, CT, KS, IL, TX, WI, GA, NC, MD, OR, AK, NH, IA, ND, OK, NM, AL, MA, KY, WA, AR, CO, IN, ME, TN, WY, NJ, SD, MS, MO, MT, DE, ID, NE, WV, RI, VT, NV, HI, DC, PR.
States: Alaska Alabama Arkansas Arizona California Colorado Connecticut District of Columbia Delaware Florida Georgia Hawaii Iowa Idaho Illinois Indiana Kansas Kentucky Louisiana Massachusetts Maryland Maine Michigan Minnesota Missouri Mississippi Montana North Carolina North Dakota Nebraska New Hampshire New Jersey New Mexico Nevada New York Ohio Oklahoma Oregon Pennsylvania Puerto Rico Rhode Island South Carolina South Dakota Tennessee Texas Utah Virginia Vermont Washington Wisconsin West Virginia Wyoming
What does Class I mean?
Class I is the most serious FDA recall class: there is a reasonable probability that using or being exposed to the product will cause serious adverse health consequences or death.
Source: FDA (medical devices), recall Z-2394-2026. Official notice. Republished from public data; RecallFlag is not affiliated with the FDA.