FDA Medical devices recall ·

Medline Kits containing Codman Disposable Perforator 14 mm: 1… recall

This is an FDA Class I recall — the most serious type.

Recall number
Z-1811-2025
Date
(initiated April 22, 2025)
Company / brand
Medline Industries, LP
3 Lakes Dr, Northfield, IL, United States
Category
Medical devices › Class I
Classification
Class I
Status
Ongoing
Units affected
1,853 total
Initiated by
Voluntary: Firm initiated

Read the official FDA notice ↗

Product

Medline Kits containing Codman Disposable Perforator 14 mm: 1. CRANI ACCESSORY PACK SKU DYNJ59270 2. CRANI PACK-LF SKU DYNJ44805M 3. CRANIOTOMY SKU CDS982719V CDS982719W DYNJ904168F DYNJ908404B DYNJ908723A DYNJ908723C 4. CRANIOTOMY #IMF 56646-LF SKU DYNJVB91001A 5. CRANIOTOMY CDS SKU CDS981888X CDS983467I 6. CRANIOTOMY PACK-LF SKU DYNJ09882G DYNJ09882I 7. CRANIOTOMY S NASSAU SKU DYNJ905879G 8. MHC CRANIOTOMY SKU DYNJ903713K DYNJ903713L 9. PAD CRANIOTOMY SKU DYNJ902149J

Why was it recalled?

Affected kits contain recalled Integra Lifesciences Codman Disposable Perforators due to inadequate welds on specific lots. Perforators with inadequate welds may break or separate.

What should you do?

FDA enforcement reports do not list a consumer remedy. Check the product codes below against what you have; if it matches, stop using it and follow the instructions in the firm's or FDA's official notice, or contact the company. If you have health concerns, contact your healthcare provider.

Product codes & lot numbers

1. CRANI ACCESSORY PACK SKU DYNJ59270 UDI-DI 10888277771512 (ea), 40888277771513 (case) Lots 23JDB683 23JDC548 24ADB783 24HDA822 2. CRANI PACK-LF SKU DYNJ44805M UDI-DI 10195327506872 (ea), 40195327506873 (case) Lots 23JBI747 23KBL228 24ABG696 24CBM117 24EBQ114 24GBJ699 3. CRANIOTOMY SKU CDS982719V UDI-DI 10195327510831 (ea), 40195327510832 (case) Lots 23KBP461 23LBN381 24ABB119 24BBC547 24BBI145 SKU CDS982719W UDI-DI 10195327645090 (ea), 40195327645091 (case) Lots 24DBS813 SKU DYNJ904168F UDI-DI 10195327395728 (ea), 40195327395729 (case) Lots 24DBU188 SKU DYNJ908404B UDI-DI 10195327410506 (ea), 40195327410507 (case) Lots 23IDB680 23LDB546 24BDB174 24FDA741 24FDB584 24IDA225 24IDA646 24IDA840 SKU DYNJ908723A UDI-DI 10195327403270 (ea), 40195327403271 (case) Lots 23IBO314 23JBC218 24ABK090 SKU DYNJ908723C UDI-DI 10198459009785 (ea), 40198459009786 (case) Lots 24IBM205 24IBN365 25ABM356 4. CRANIOTOMY #IMF 56646-LF SKU DYNJVB91001A UDI-DI 10195327466466 (ea), 40195327466467 (case) Lots 23JLA765 24DLA724 24DLB006 24ELA325 5. CRANIOTOMY CDS SKU CDS981888X UDI-DI 10195327144760 (ea), 40195327144761 (case) Lots 24ABT432 24CBD698 24HBH954 24HBP483 24IBG797 25CBF098 SKU CDS983467I UDI-DI 10195327457945 (ea), 40195327457946 (case) Lots 23KDA336 6. CRANIOTOMY PACK-LF SKU DYNJ09882G UDI-DI 10195327574055 (ea), 40195327574056 (case) Lots 24CBF778 SKU DYNJ09882I UDI-DI 10198459058172 (ea), 40198459058173 (case) Lots 24HBN426 7. CRANIOTOMY S NASSAU SKU DYNJ905879G UDI-DI 10195327512996 (ea),

Distribution

US distribution to CA, FL, IL, IN, KY, MA, MD, MN, NY, TX, VT, WA and WI. No OUS distribution.

States: California Florida Illinois Indiana Kentucky Massachusetts Maryland Minnesota New York Texas Vermont Washington Wisconsin

What does Class I mean?

Class I is the most serious FDA recall class: there is a reasonable probability that using or being exposed to the product will cause serious adverse health consequences or death.

Source: FDA (medical devices), recall Z-1811-2025. Official notice. Republished from public data; RecallFlag is not affiliated with the FDA.