FDA Medical devices recall ·

Medline Kits containing Codman Disposable Perforator 14 mm: 1… recall

This is an FDA Class I recall — the most serious type.

Recall number
Z-1809-2025
Date
(initiated April 22, 2025)
Company / brand
Medline Industries, LP
3 Lakes Dr, Northfield, IL, United States
Category
Medical devices › Class I
Classification
Class I
Status
Ongoing
Units affected
1853 total
Initiated by
Voluntary: Firm initiated

Read the official FDA notice ↗

Product

Medline Kits containing Codman Disposable Perforator 14 mm: 1. CRANI ADD A PACK, SKU DYNJ30926O 2. CRANIOTOMY, SKU CDS983611N 3. CRANIOTOMY CDS, SKU CDS983467J 4. CRANIOTOMY PACK, SKU DYNJ85807A, DYNJ85927 5. MAJOR CRANIOTOMY PACK, SKU DYNJ82007A

Why was it recalled?

Affected kits contain recalled Integra Lifesciences Codman Disposable Perforators due to inadequate welds on specific lots. Perforators with inadequate welds may break or separate.

What should you do?

FDA enforcement reports do not list a consumer remedy. Check the product codes below against what you have; if it matches, stop using it and follow the instructions in the firm's or FDA's official notice, or contact the company. If you have health concerns, contact your healthcare provider.

Product codes & lot numbers

1. CRANI ADD A PACK, SKU DYNJ30926O, UDI-DI 10195327539153 (EA), 40195327539154 (CASE) KIT LOTS 24AMF811 2. CRANIOTOMY, SKU CDS983611N, UDI-DI 10195327466626 (EA), 40195327466627 (CASE) KIT LOTS 23KBA720 23LBI611 23LBJ586 3. CRANIOTOMY CDS, SKU CDS983467J, UDI-DI 10195327561116 (EA), 40195327561117 (CASE) KIT LOTS 24ADA818 24ADC216 24BDA929 24DDB447 24EDA275 24EDA751 24GDB213 24IDA327 24IDB043 4. CRANIOTOMY PACK, SKU DYNJ85807A, UDI-DI 10195327637392 (EA), 40195327637393 (CASE) KIT LOTS 24CBB843 24EBC889 24GBY029 ; SKU DYNJ85927, UDI-DI 10195327575571 (EA), 40195327575572 (CASE) KIT LOTS 24BMD310 24EMI836 5. MAJOR CRANIOTOMY PACK, SKU DYNJ82007A UDI-DI 10195327635466 (EA), 40195327635467 (CASE) KIT LOTS 24CMA496 24HMI081 24HMI978

Distribution

US distribution to CA, FL, IL, IN, KY, MA, MD, MN, NY, TX, VT, WA and WI. No OUS distribution.

States: California Florida Illinois Indiana Kentucky Massachusetts Maryland Minnesota New York Texas Vermont Washington Wisconsin

What does Class I mean?

Class I is the most serious FDA recall class: there is a reasonable probability that using or being exposed to the product will cause serious adverse health consequences or death.

Source: FDA (medical devices), recall Z-1809-2025. Official notice. Republished from public data; RecallFlag is not affiliated with the FDA.