FDA Medical devices recall ·

Medline Convenience Kits: 1) MAJOR LAPAROTOMY CDS, Model Number: CDS860015U recall

Recall number
Z-1395-2026
Date
(initiated January 7, 2026)
Company / brand
Medline Industries, LP
3 Lakes Dr, Northfield, IL, United States
Category
Medical devices › Class II
Classification
Class II
Status
Ongoing
Units affected
82597 units
Hazard tags
Sterility
Initiated by
Voluntary: Firm initiated

Read the official FDA notice ↗

Product

Medline Convenience Kits: 1) MAJOR LAPAROTOMY CDS, Model Number: CDS860015U; 2) MAJOR LAPAROTOMY CDS, Model Number: CDS860015V; 3) MAJOR LAPAROTOMY CDS, Model Number: CDS860015W; 4) MAJOR LAPAROTOMY CDS, Model Number: CDS860015X; 5) MINOR LAPAROTOMY CDS, Model Number: CDS860016P; 6) MINOR LAPAROTOMY CDS, Model Number: CDS860016Q; 7) MINOR LAPAROTOMY CDS, Model Number: CDS860016R; 8) MINOR LAPAROTOMY CDS, Model Number: CDS860016S; 9) LAP GASTRIC BYPASS CDS, Model Number: CDS860146M; 10) NEW LONDON MINOR CDS, Model Number: CDS982518M; 11) NEW LONDON MINOR CDS, Model Number: CDS982518N; 12) NEW LONDON MINOR CDS, Model Number: CDS982518O; 13) NEW LONDON MINOR CDS, Model Number: CDS982518P; 14) GENERAL LAPAROSCOPY CDS, Model Number: CDS983311J; 15) LAVH CDS, Model Number: CDS983411G; 16) NHP ROBOTIC GENERAL CDS, Model Number: CDS984244J; 17) ROBOTIC-LF, Model Number: CDS984543C; 18) CYSTO CDS, Model Number: CDS984592G; 19) CYSTO CDS, Model Number: CDS984592I; 20) ROBOTIC, Model Number: CDS985270F; 21) GENERAL LAPAROSCOPY CDS, Model Number: CDS985422I; 22) ROBOTIC CDS, Model Number: CDS985429L; 23) LAP CHOLE CDS, Model Number: CDS985558N; 24) LAP CHOLE CDS, Model Number: CDS985558O; 25) ENDO KIT, Model Number: DYKE1580A; 26) KIT GEN SURG EXLAP, Model Number: DYKMBNDL105; 27) KIT GEN SURG EXLAP, Model Number: DYKMBNDL105A; 28) KIT GEN SURG EXLAP, Model Number: DYKMBNDL105B; 29) KIT ROBOTICS UROLOGY PROSTATE, Model Number: DYKMBNDL116; 30) KIT ROBOTICS UROLOGY PROSTATE, Model Number: DYKMBNDL116B; 31) KIT UROLOGY URETHROPLASTY, Model Number: DYKMBNDL117A; 32) KIT SURG ONC DIAGNOSTIC LAPARO, Model Number: DYKMBNDL12; 33) KIT UROLOGY LAP PHELAN, Model Number: DYKMBNDL121; 34) KIT UROLOGY LAP PHELAN, Model Number: DYKMBNDL121A; 35) KIT UROLOGY LAP PHELAN, Model Number: DYKMBNDL121B; 36) KIT UROLOGY LAP PHELAN, Model Number: DYKMBNDL121C; 37) KIT UROLOGY LAP PHELAN, Model Number: DYKMBNDL121F; 38) KIT L&D CYSTO SETUP, Model Number: DYKMBNDL135A; 39) KIT THORACIC LAPAROSCOPIC HIAT, Model Number: DYKMBNDL136; 40) KIT THORACIC LAPAROSCOPIC HIAT, Model Number: DYKMBNDL136A; 41) KIT THORACIC STAGING LAPAROTOM, Model Number: DYKMBNDL15; 42) KIT SURG ONC LAP DIAG/CV CATH, Model Number: DYKMBNDL153A; 43) KIT SURG ONC LAP DIAG/PORT INS, Model Number: DYKMBNDL153B; 44) KIT LAP DONOR NEPHRECTOMY, Model Number: DYKMBNDL173; 45) KIT LAP DONOR NEPHRECTOMY, Model Number: DYKMBNDL173A; 46) KIT THORACIC HIATAL HERNIA TRA, Model Number: DYKMBNDL175; 47) KIT GEN SURG LAP INGUINAL, Model Number: DYKMBNDL176; 48) KIT GEN SURG LAP INGUINAL, Model Number: DYKMBNDL176A; 49) KIT GEN SURG LAP INGUINAL, Model Number: DYKMBNDL176B; 50) KIT GEN SURG LAP INGUINAL, Model Number: DYKMBNDL176C; 51) KIT THORACIC ROBOTIC LOBECTOMY, Model Number: DYKMBNDL184; 52) KIT THORACIC ROBOTIC LOBECTOMY, Model Number: DYKMBNDL184A; 53) KIT UROLOGY PENILE REPAIR, Model Number: DYKMBNDL188A; 54) KIT UROLOGY LAP SIDDIQUI, Model Number: DYKMBNDL193; 55) KIT UROLOGY LAP SIDDIQUI, Model Number: DYKMBNDL193A; 56) KIT UROLOGY BUCCAL GRAFT, Model Number: DYKMBNDL206; 57) KIT STC PLASTIC FREE FLAP, Model Number: DYKMBNDL54A; 58) KIT UROLOGY CYSTOSCOPY SUPPLEM, Model Number: DYKMBNDL60; 59) KIT UROLOGY CYSTOSCOPY SUPPLEM, Model Number: DYKMBNDL60A; 60) KIT UROLOGY MINOR, Model Number: DYKMBNDL61A; 61) KIT UROLOGY MINOR, Model Number: DYKMBNDL61B; 62) KIT GEN SURG LAP COLECTOMY, Model Number: DYKMBNDL73; 63) KIT GEN SURG LAP COLORECTAL, Model Number: DYKMBNDL73A; 64) KIT GEN SURG LAP COLORECTAL, Model Number: DYKMBNDL73B; 65) KIT GEN SURG LAP COLORECTAL, Model Number: DYKMBNDL73C; 66) KIT GEN SURG LAP COLORECTAL, Model Number: DYKMBNDL73D; 67) KIT GEN SURG LAP GASTRIC BYPAS, Model Number: DYKMBNDL74; 68) KIT GEN SURG LAP GASTRIC BYPAS, Model Number: DYKMBNDL74A; 69) KIT GEN SURG LAP GASTRIC BYPAS, Model Number: DYKMBNDL74B; 70) KIT GEN SURG LAP GASTRIC BYPAS,

Why was it recalled?

Medline has identified issues related to calibration of the equipment used to sterilize and package the devices. All products were exposed to the validated sterilization and packaging cycles; however, the identified calibration issues have the potential to impact the sterility assurance level (SAL) of the Recalled Products.

What should you do?

FDA enforcement reports do not list a consumer remedy. Check the product codes below against what you have; if it matches, stop using it and follow the instructions in the firm's or FDA's official notice, or contact the company. If you have health concerns, contact your healthcare provider.

Product codes & lot numbers

1) CDS860015U, UDI-DI: 10193489847994(each), 40193489847995(case), Lot Number: 21LBT172; 2) CDS860015U, UDI-DI: 10193489847994(each), 40193489847995(case), Lot Number: 21LBD726; 3) CDS860015U, UDI-DI: 10193489847994(each), 40193489847995(case), Lot Number: 21KBB595; 4) CDS860015U, UDI-DI: 10193489847994(each), 40193489847995(case), Lot Number: 21IBL031; 5) CDS860015U, UDI-DI: 10193489847994(each), 40193489847995(case), Lot Number: 21GBS688; 6) CDS860015U, UDI-DI: 10193489847994(each), 40193489847995(case), Lot Number: 21FBR386; 7) CDS860015V, UDI-DI: 10195327209582(each), 40195327209583(case), Lot Number: 22HBF170; 8) CDS860015W, UDI-DI: 10195327349509(each), 40195327349500(case), Lot Number: 23LBQ388; 9) CDS860015W, UDI-DI: 10195327349509(each), 40195327349500(case), Lot Number: 23KBX194; 10) CDS860015W, UDI-DI: 10195327349509(each), 40195327349500(case), Lot Number: 23JBR469; 11) CDS860015W, UDI-DI: 10195327349509(each), 40195327349500(case), Lot Number: 23HBT489; 12) CDS860015W, UDI-DI: 10195327349509(each), 40195327349500(case), Lot Number: 23GBM065; 13) CDS860015W, UDI-DI: 10195327349509(each), 40195327349500(case), Lot Number: 23FBP597; 14) CDS860015X, UDI-DI: 10195327647513(each), 40195327647514(case), Lot Number: 25EBQ940; 15) CDS860015X, UDI-DI: 10195327647513(each), 40195327647514(case), Lot Number: 25DBF532; 16) CDS860015X, UDI-DI: 10195327647513(each), 40195327647514(case), Lot Number: 24EBA340; 17) CDS860015X, UDI-DI: 10195327647513(each), 40195327647514(case), L

Distribution

Worldwide distribution - US Nationwide.

What does Class II mean?

Class II means use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or the probability of serious harm is remote.