FDA Medical devices recall ·

Medline Convenience Kits: 1) SCC RF PAIN MGMT, Model Number: DYNJ56436A recall

Recall number
Z-1386-2026
Date
(initiated January 7, 2026)
Company / brand
Medline Industries, LP
3 Lakes Dr, Northfield, IL, United States
Category
Medical devices › Class II
Classification
Class II
Status
Ongoing
Units affected
779 units
Hazard tags
Sterility
Initiated by
Voluntary: Firm initiated

Read the official FDA notice ↗

Product

Medline Convenience Kits: 1) SCC RF PAIN MGMT, Model Number: DYNJ56436A; 2) SCC RF PAIN MGMT, Model Number: DYNJ56436B; 3) CPNB SETUP TRAY, Model Number: DYNJRA1979; 4) CPNB SETUP TRAY, Model Number: DYNJRA1979A; 5) CPNB SETUP TRAY, Model Number: SAMPA0108

Why was it recalled?

Medline has identified issues related to calibration of the equipment used to sterilize and package the devices. All products were exposed to the validated sterilization and packaging cycles; however, the identified calibration issues have the potential to impact the sterility assurance level (SAL) of the Recalled Products.

What should you do?

FDA enforcement reports do not list a consumer remedy. Check the product codes below against what you have; if it matches, stop using it and follow the instructions in the firm's or FDA's official notice, or contact the company. If you have health concerns, contact your healthcare provider.

Product codes & lot numbers

1) DYNJ56436A, UDI-DI: 10889942749874(each), 40889942749875(case), Lot Number: 21BMC737; 2) DYNJ56436A, UDI-DI: 10889942749874(each), 40889942749875(case), Lot Number: 21AMB799; 3) DYNJ56436B, UDI-DI: 10193489869569(each), 40193489869560(case), Lot Number: 21DME244; 4) DYNJRA1979, UDI-DI: 10195327186852(each), 40195327186853(case), Lot Number: 22IBU628; 5) DYNJRA1979, UDI-DI: 10195327186852(each), 40195327186853(case), Lot Number: 22GBV936; 6) DYNJRA1979A, UDI-DI: 10195327649258(each), 40195327649259(case), Lot Number: 25DME153; 7) DYNJRA1979A, UDI-DI: 10195327649258(each), 40195327649259(case), Lot Number: 24KMA261; 8) DYNJRA1979A, UDI-DI: 10195327649258(each), 40195327649259(case), Lot Number: 24JMA460; 9) SAMPA0108, UDI-DI: 10195327096861(each), 40195327096862(case), Lot Number: 22NBG086; 10) SAMPA0108, UDI-DI: 10195327096861(each), 40195327096862(case), Lot Number: 22BBD381

Distribution

Worldwide distribution - US Nationwide.

What does Class II mean?

Class II means use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or the probability of serious harm is remote.

Source: FDA (medical devices), recall Z-1386-2026. Official notice. Republished from public data; RecallFlag is not affiliated with the FDA.