FDA Medical devices recall ·
Medline Convenience Kits: 1) ANCILLARY PACK-LF, Model Number: DYNJ0185290G recall
- Recall number
- Z-1384-2026
- Date
- (initiated January 7, 2026)
- Company / brand
- Medline Industries, LP
3 Lakes Dr, Northfield, IL, United States - Category
- Medical devices › Class II
- Classification
- Class II
- Status
- Ongoing
- Units affected
- 551 units
- Hazard tags
- Sterility
- Initiated by
- Voluntary: Firm initiated
Read the official FDA notice ↗
Product
Medline Convenience Kits: 1) ANCILLARY PACK-LF, Model Number: DYNJ0185290G; 2) OPEN HEART ANESTHESIA ICC-A, Model Number: DYNJ905567A
Why was it recalled?
Medline has identified issues related to calibration of the equipment used to sterilize and package the devices. All products were exposed to the validated sterilization and packaging cycles; however, the identified calibration issues have the potential to impact the sterility assurance level (SAL) of the Recalled Products.
What should you do?
FDA enforcement reports do not list a consumer remedy. Check the product codes below against what you have; if it matches, stop using it and follow the instructions in the firm's or FDA's official notice, or contact the company. If you have health concerns, contact your healthcare provider.
Product codes & lot numbers
1) DYNJ0185290G, UDI-DI: 10193489378054(each), 40193489378055(case), Lot Number: 22LBQ336; 2) DYNJ905567A, UDI-DI: 10193489348927(each), 40193489348928(case), Lot Number: 24BBN429; 3) DYNJ905567A, UDI-DI: 10193489348927(each), 40193489348928(case), Lot Number: 23KBI089; 4) DYNJ905567A, UDI-DI: 10193489348927(each), 40193489348928(case), Lot Number: 23HBG596; 5) DYNJ905567A, UDI-DI: 10193489348927(each), 40193489348928(case), Lot Number: 23FBL644; 6) DYNJ905567A, UDI-DI: 10193489348927(each), 40193489348928(case), Lot Number: 23FBG152; 7) DYNJ905567A, UDI-DI: 10193489348927(each), 40193489348928(case), Lot Number: 23EBH550; 8) DYNJ905567A, UDI-DI: 10193489348927(each), 40193489348928(case), Lot Number: 23CBQ546; 9) DYNJ905567A, UDI-DI: 10193489348927(each), 40193489348928(case), Lot Number: 22NBF780; 10) DYNJ905567A, UDI-DI: 10193489348927(each), 40193489348928(case), Lot Number: 22LBC300; 11) DYNJ905567A, UDI-DI: 10193489348927(each), 40193489348928(case), Lot Number: 22HBP453; 12) DYNJ905567A, UDI-DI: 10193489348927(each), 40193489348928(case), Lot Number: 22HBM263; 13) DYNJ905567A, UDI-DI: 10193489348927(each), 40193489348928(case), Lot Number: 22HBF686; 14) DYNJ905567A, UDI-DI: 10193489348927(each), 40193489348928(case), Lot Number: 22CBY841; 15) DYNJ905567A, UDI-DI: 10193489348927(each), 40193489348928(case), Lot Number: 21KBR457; 16) DYNJ905567A, UDI-DI: 10193489348927(each), 40193489348928(case), Lot Number: 21KBA253; 17) DYNJ905567A, UDI-DI: 10193489348927(each), 40193
Distribution
Worldwide distribution - US Nationwide.
What does Class II mean?
Class II means use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or the probability of serious harm is remote.
Source: FDA (medical devices), recall Z-1384-2026. Official notice. Republished from public data; RecallFlag is not affiliated with the FDA.