FDA Medical devices recall ·
Medline custom medical procedure kits labeled as: 1) VP SHUNT CDS recall
- Recall number
- Z-0779-2025
- Date
- (initiated November 21, 2024)
- Company / brand
- Medline Industries, LP
3 Lakes Dr, Northfield, IL, United States - Category
- Medical devices › Class II
- Classification
- Class II
- Status
- Ongoing
- Units affected
- 1402 units
- Hazard tags
- Sterility Microbial contamination
- Initiated by
- Voluntary: Firm initiated
Read the official FDA notice ↗
Product
Medline custom medical procedure kits labeled as: 1) VP SHUNT CDS, REF CDS983470D; 2) 1/2X1/2 NEURO SPONGE PACK-LF, REF DYNJ0713177; 3) BURR HOLE SHUNT PACK, REF DYNJ56819D; 4) BURR HOLE SHUNT PACK, REF DYNJ56819F; 5) BURR HOLE SHUNT PACK, REF DYNJ56819G; 6) RO VP SHUNT PACK, REF PHS392834010D; 7) VP SHUNT, REF SYNJ10307A.
Why was it recalled?
Custom convenience kits are being recall because the contain non-sterile Codman Surgical Patties & Surgical Strips, which have the potential for endotoxin contamination.
What should you do?
FDA enforcement reports do not list a consumer remedy. Check the product codes below against what you have; if it matches, stop using it and follow the instructions in the firm's or FDA's official notice, or contact the company. If you have health concerns, contact your healthcare provider.
Product codes & lot numbers
1) REF CDS983470D, UDI/DI 10193489275339 (EA) 40193489275330(CS), Lot numbers: 20BDB519, 20EDA360, 20EDC040, 20GDA474, 20JDA107, 20KDA855, 21CDA110, 21EDA674, 21FDA748; 2) REF DYNJ0713177, UDI/DI 10889942085088 (EA) 40889942085089(CS), Lot numbers: 19AKC072, 19HKC645, 19KKA304, 20FKA539; 3) REF DYNJ56819D, UDI/DI 10193489254273 (EA) 40193489254274(CS), Lot numbers: 20AKA332, 20BKA860; 4) REF DYNJ56819F, UDI/DI 10193489353532 (EA) 40193489353533(CS), Lot numbers: 20EKB648, 20HKA426, 20JKA217, 20LKB026, 21FBT545, 21HBL828, 21KBY910, 21KBY914; 5) REF DYNJ56819G, UDI/DI 10195327073749 (EA) 40195327073740(CS), Lot numbers: 22CBJ951, 22HBJ403, 22LBQ584, 23BBH164, 23CBG816, 23FBM524; 6) REF PHS392834010D, UDI/DI 10193489222432 (EA) 40193489222433(CS), Lot numbers: 21LDA084; 7) REF SYNJ10307A, UDI/DI 10193489377026 (EA) 40193489377027(CS), Lot numbers: 20HBU444, 20HBW312, 21EBD895, 21FBE508, 21HBS523.
Distribution
Worldwide distribution - US Nationwide and the countries of UAE, and Canada.
What does Class II mean?
Class II means use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or the probability of serious harm is remote.
Source: FDA (medical devices), recall Z-0779-2025. Official notice. Republished from public data; RecallFlag is not affiliated with the FDA.