FDA Medical devices recall ·

Medline custom medical procedure kits labeled as: 1) OPEN HEART CDS-LF… recall

Recall number
Z-0778-2025
Date
(initiated November 21, 2024)
Company / brand
Medline Industries, LP
3 Lakes Dr, Northfield, IL, United States
Category
Medical devices › Class II
Classification
Class II
Status
Ongoing
Units affected
2344 units
Hazard tags
Sterility Microbial contamination
Initiated by
Voluntary: Firm initiated

Read the official FDA notice ↗

Product

Medline custom medical procedure kits labeled as: 1) OPEN HEART CDS-LF, REFCDS840451M; 2) NEURO PACK, REFDYNJ41516M; 3) RFD- AAA PACK, REFDYNJ54242B; 4) OPEN HEART A&B, REFDYNJ901203C; 5) OPEN HEART A&B, REFDYNJ901203D; 6) ACDF-LF, REFDYNJ906681B; 7) ACDF-LF, REFDYNJ906681C.

Why was it recalled?

Custom convenience kits are being recall because the contain non-sterile Codman Surgical Patties & Surgical Strips, which have the potential for endotoxin contamination.

What should you do?

FDA enforcement reports do not list a consumer remedy. Check the product codes below against what you have; if it matches, stop using it and follow the instructions in the firm's or FDA's official notice, or contact the company. If you have health concerns, contact your healthcare provider.

Product codes & lot numbers

1) REF CDS840451M, UDI/DI 10195327417574 (EA) 40195327417575 (CS), Lot Numbers: 24BLA736; 2) REF DYNJ41516M, UDI/DI 10193489202564 (EA) 40193489202565 (CS), Lot Numbers: 19JKB417, 19LKA454, 20AKB109, 20DKB147, 20FKB171; 3) REF DYNJ54242B, UDI/DI 10195327167097 (EA) 40195327167098 (CS), Lot Numbers: 23GBO400; 4) REF DYNJ901203C, UDI/DI 10889942572526 (EA) 40889942572527 (CS), Lot numbers: 19BKA270, 19CKC053, 19EKD526, 19FKC478, 19HKA796, 19KKA783, 19LKB868, 20BKA558, 20FKA993, 20IKA397, 20LKA831, 21BKA583, 21DKA592, 21GMB734; 5) REF DYNJ901203D, UDI/DI 10193489947526 (EA) 40193489947527 (CS), Lot Numbers: 21GME287, 21JMH613, 22AMB438, 22CMD928, 22CMF425, 22EMF242, 22FMG716, 22GMH022, 22IMD195, 23AMB881, 23DME044, 23DME245, 23EMI884, 23HMH269, 23HMI447, 23LMB223, 24AMB281, 24CMI324, 24EMC691, 24EMI830; 6) REF DYNJ906681B, UDI/DI 10195327106881 (EA) 40195327106882 (CS), Lot Numbers: 22FBC663, 22OBM278; 7) REF DYNJ906681C, UDI/DI 10195327173838 (EA) 40195327173839 (CS) , Lot Numbers: 22IBR002, 22JBV571, 22JBW937, 22KBB923, 22LBK562, 23ABC493, 23GBR644, 23IBP138, 23IBV201, 23JBK264, 23LBE604, 23LBS964, 24CBP708, 24EBA703, 24GBL458.

Distribution

Worldwide distribution - US Nationwide and the countries of UAE, and Canada.

What does Class II mean?

Class II means use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or the probability of serious harm is remote.

Source: FDA (medical devices), recall Z-0778-2025. Official notice. Republished from public data; RecallFlag is not affiliated with the FDA.