FDA Medical devices recall ·

Generic Name: restor3d MTP Implant Generic Name: MTP Implant Model recall

Recall number
Z-0264-2027
Date
(initiated September 4, 2026)
Company / brand
restor3d Inc.
4001 E Nc 54 Hwy, Durham, NC, United States
Category
Medical devices › Class II
Classification
Class II
Status
Ongoing
Units affected
16
Hazard tags
Sterility
Initiated by
Voluntary: Firm initiated

Read the official FDA notice ↗

Product

Generic Name: restor3d MTP Implant Generic Name: MTP Implant Model/Catalog Number: 1103-161620 Software Version: None Product Description: The restor3d Metatarsophalangeal (MTP) Implant is a sterile, single-use implant grade metal device, available in varied articulating surface coverages. The restor3d MTP implant is designed to replace the damaged articulating surface of the first metatarsal distal head. The device is intended for the treatment of patients with degenerative and post-traumatic arthritis in the first metatarsal joint in the presence of good bone stock and integrity of the first metatarsal head. This implant can also be used in the following clinical conditions: hallux valgus or hallux limitus, hallux rigidus, and an unstable or painful metatarsal/phalangeal (MTP) joint. The implant is a single use product intended to be press fit, with optional use of bone cement. The restor3d MTP Implant is comprised of a single laser powder bed fusion (LPBF) printed cobalt-chromium alloy (Co28Cr6Mo) part. The external surface of the implant that is intended to be in contact with the phalanx is mirror polished. Component: Not a component

Why was it recalled?

Implants were found in the field with orange sterilzation indicators which means implants were not exposed to sterilization.

What should you do?

FDA enforcement reports do not list a consumer remedy. Check the product codes below against what you have; if it matches, stop using it and follow the instructions in the firm's or FDA's official notice, or contact the company. If you have health concerns, contact your healthcare provider.

Product codes & lot numbers

Lot Code: Model: 1103-161620 UDI: 00840097505244 Lot: A02979

Distribution

U.S. Nationwide distribution in the states of FL, IL, NC, PA, SC, and TX.

States: Florida Illinois North Carolina Pennsylvania South Carolina Texas

What does Class II mean?

Class II means use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or the probability of serious harm is remote.

Source: FDA (medical devices), recall Z-0264-2027. Official notice. Republished from public data; RecallFlag is not affiliated with the FDA.