FDA Medical devices recall ·

Veritas Reverse Total Shoulder Replacement, Model recall

Recall number
Z-0348-2027
Date
(initiated October 1, 2026)
Company / brand
restor3d Inc.
4001 E Nc 54 Hwy, Durham, NC, United States
Category
Medical devices › Class II
Classification
Class II
Status
Ongoing
Units affected
240 units
Initiated by
Voluntary: Firm initiated

Read the official FDA notice ↗

Product

Veritas Reverse Total Shoulder Replacement, Model/Catalog Number: 3500-140C3, 3500-140U3, 3500-140C1, 3500-140U1, size 1 -2, 40mm Polymer Liner

Why was it recalled?

(Retroactively reported) Pairing of shoulder implant stem sizes 1 and 2 with the largest glenosphere diameter (40 mm) creates an excessive mechanical loading configuration in which certain polyethylene components lack adequate retention strength, potentially resulting in dislocation, reduced range of motion, or revision surgery.

What should you do?

FDA enforcement reports do not list a consumer remedy. Check the product codes below against what you have; if it matches, stop using it and follow the instructions in the firm's or FDA's official notice, or contact the company. If you have health concerns, contact your healthcare provider.

Product codes & lot numbers

All Lot codes/ UDIs:Product Code: 3500-140C3 UDI/GTIN:00840097513492; Product Code: 3500-140U3 UDI/GTIN:00840097513461; Product Code: 3500-140C1 UDI/GTIN:00840097513430; Product Code: 3500-140U1 UDI/GTIN:00840097513409

Distribution

US Nationwide distribution in the states of Texas, Massachusetts, Oklahoma, Florida, Colorado, North Carolina.

States: Colorado Florida Massachusetts North Carolina Oklahoma Texas

What does Class II mean?

Class II means use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or the probability of serious harm is remote.

Source: FDA (medical devices), recall Z-0348-2027. Official notice. Republished from public data; RecallFlag is not affiliated with the FDA.