FDA Medical devices recall ·

Medical convenience kits Item Number/Description EVLH67F LEFT HEART CATH… recall

Recall number
Z-0175-2026
Date
(initiated September 4, 2025)
Company / brand
American Contract Systems Inc
1601 Se Gateway Dr Ste 120, Grimes, IA, United States
Category
Medical devices › Class II
Classification
Class II
Status
Ongoing
Units affected
32,433 total
Hazard tags
Sterility
Initiated by
Voluntary: Firm initiated

Read the official FDA notice ↗

Product

Medical convenience kits Item Number/Description EVLH67F LEFT HEART CATH PACK FDCL22R CATH LAB ANGIO IHCC03X CATH CARDIAC PACK TNCC02AD CARDIAC CATH TRINITY UICC27AG CARDIAC CATH ANGIO PACK (PS 923077) UIRD89AH RADIAL PACK (PS 56573)

Why was it recalled?

Identified products were re-gassed (sterilized more than once) after a nonconformance occurred during the initial Ethylene Oxide (EO) gas injection process. These products have not been validated for exposure to multiple sterilization cycles. Therefore, product quality, safety, and effectiveness cannot be confirmed for product that was re-gassed.

What should you do?

FDA enforcement reports do not list a consumer remedy. Check the product codes below against what you have; if it matches, stop using it and follow the instructions in the firm's or FDA's official notice, or contact the company. If you have health concerns, contact your healthcare provider.

Product codes & lot numbers

AMCL05P UDI-DI 191072207296 Lot 7990011 Bag serial no. N/A Lot 8170111 Bag serial numbers 68194152 68194153 68194160 68194176 68194179 68194182 68194190 Lot 8526511 Bag serial numbers 68873157 68874410; EVLH67F UDI-DI 191072232380 Lot 8211011 Bag serial number N/A Lot 8367311 Bag serial number 68570438; FDCL22R UDI-DI 191072185259 Lot 8048511 Bag serial numbers 68045373 68045376; IHCC03X UDI-DI 191072226273 Lot 80111 Bag serial number N/A Lot 8085711 Bag serial numbers 47506640 47506761; TNCC02AD UDI-DI 191072212757 Lot 7970911 Bag serial number N/A Lot 8045911 Bag serial number 47503192 Lot 8426711 Bag serial number 47886615; UICC27AG UDI-DI 191072231307 Lot 8085511 Bag serial number N/A Lot 8167611 Bag serial number 68194108; UIRD89AH UDI-DI 191072231314 Lot 8068111 Bag serial number 47499672 47499908 * if Bag serial number is "n/a", then all of that lot is affected

Distribution

US Nationwide distribution in the states of SD, IA, MN, WA, IL.

States: Iowa Illinois Minnesota South Dakota Washington

What does Class II mean?

Class II means use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or the probability of serious harm is remote.

Source: FDA (medical devices), recall Z-0175-2026. Official notice. Republished from public data; RecallFlag is not affiliated with the FDA.