FDA Medical devices recall ·
Medline surgical convenience kits labeled as: CABG AUXILIARY PACK… recall
- Recall number
- Z-0168-2027
- Date
- (initiated September 2, 2026)
- Company / brand
- Medline Industries, LP
3 Lakes Dr, Northfield, IL, United States - Category
- Medical devices › Class II
- Classification
- Class II
- Status
- Ongoing
- Units affected
- 3125 kits
- Initiated by
- Voluntary: Firm initiated
Read the official FDA notice ↗
Product
Medline surgical convenience kits labeled as: CABG AUXILIARY PACK, Medline Product Number/SKU DYNJ44925D
Why was it recalled?
Medical convenience kits were distributed containing cannula components which do not conform to ISO 80369-7 dimensional requirements and may contribute to leaking or breaking during use.
What should you do?
FDA enforcement reports do not list a consumer remedy. Check the product codes below against what you have; if it matches, stop using it and follow the instructions in the firm's or FDA's official notice, or contact the company. If you have health concerns, contact your healthcare provider.
Product codes & lot numbers
UDI/DI 10193489461725 (each) 40193489461726 (case), Lot Numbers: 25LMF813, 25IMF341, 25HBR484, 25FBP179, 25EBJ732, 25DBK745, 25CBK191, 25BBE801, 24LBB945, 24IBT658, 24IBD265, 24GBG100, 24EBC752, 24CBQ893, 24BBJ088, 24ABE008, 23JBA374, 23HBV035, 23GBL681, 23EBK074, 23DBL164, 23DBB447, 23ABV418, 22LBG352, 22IBK243, 22HBX829, 22EBA032, 22DBO915, 22BBR573, 21LBB905, 21IBX719, 21HBG260, 21GBH209, 21DBR261, 21BBJ878.
Distribution
US Nationwide distribution.
What does Class II mean?
Class II means use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or the probability of serious harm is remote.
Source: FDA (medical devices), recall Z-0168-2027. Official notice. Republished from public data; RecallFlag is not affiliated with the FDA.