FDA Medical devices recall ·
Medline surgical convenience kits labeled as: 1) OPEN HEART PACK A & recall
- Recall number
- Z-0176-2027
- Date
- (initiated September 2, 2026)
- Company / brand
- Medline Industries, LP
3 Lakes Dr, Northfield, IL, United States - Category
- Medical devices › Class II
- Classification
- Class II
- Status
- Ongoing
- Units affected
- 36199 kits
- Initiated by
- Voluntary: Firm initiated
Read the official FDA notice ↗
Product
Medline surgical convenience kits labeled as: 1) OPEN HEART PACK A & B, Medline Product number/SKU DYNJ04584L; 2) OPEN HEART ACCESSORY PACK, Medline Product number/SKU DYNJT5715A; 3) CABG ACCESSORY PACK, Medline Product number/SKU DYNJ85751; 4) OPEN HEART PACK, Medline Product number/SKU DYNJ53741N; 5) OPEN HEART PACK 1&2, Medline Product number/SKU DYNJ80701A; 6) VEIN PACK, Medline Product number/SKU DYNJ67308C; 7) OPEN HEART PACK, Medline Product number/SKU DYNJ89451B; 8) VASCULAR TRAY-LF, Medline Product number/SKU DYNJ42166A; 9) ENDOSCOPIC VEIN HARVEST PK, Medline Product number/SKU DYNJ67363C; 10) ENDOSCOPIC VEIN HARVEST PK, Medline Product number/SKU DYNJ67363B; 11) LIVER TRANSPLANT PACK B, Medline Product number/SKU DYNJ65779B; 12) CABG PACK-LF, Medline Product number/SKU DYNJ47661J; 13) VEIN HARVEST PACK, Medline Product number/SKU DYNJ04470L; 14) CARDIAC PACK, Medline Product number/SKU DYNJ44647D; 15) GM OPEN HEART ACC PACK-LF, Medline Product number/SKU DYNJ0660029AM; 16) CARDIOVASCULAR PACK-LF, Medline Product number/SKU DYNJ0160327C; 17) GOOD SHEPHERD OPEN HEART, Medline Product number/SKU DYNJ60819L; 18) OPEN HEART PACK, Medline Product number/SKU DYNJ0271227R; 19) OPEN HEART PACK, Medline Product number/SKU DYNJ81541C; 20) TAVR PACK, Medline Product number/SKU DYNJ54094F; 21) TRANSPLANT BACKTABLE PACK, Medline Product number/SKU DYNJ62055C; 22) VASCULAR, Medline Product number/SKU DYNJ49366; 23) BACK TABLE TRAY, Medline Product number/SKU DYNJ65416D; 24) CV PART 1 AND 2, Medline Product number/SKU DYNJ908879G; 25) OPEN HEART A B PACK, Medline Product number/SKU DYNJ81096B; 26) KIDNEY, Medline Product number/SKU DYNJ54732I; 27) ABDOMINAL PRESEVERVATION PACK, Medline Product number/SKU DYNJ62691F; 28) LIVER PACK SAFETY, Medline Product number/SKU DYNJ54734I; 29) CV PROCEDURE PART A, Medline Product number/SKU DYNJ903745I;
Why was it recalled?
Medical convenience kits were distributed containing cannula components which do not conform to ISO 80369-7 dimensional requirements and may contribute to leaking or breaking during use.
What should you do?
FDA enforcement reports do not list a consumer remedy. Check the product codes below against what you have; if it matches, stop using it and follow the instructions in the firm's or FDA's official notice, or contact the company. If you have health concerns, contact your healthcare provider.
Product codes & lot numbers
Medline product number/SKU DYNJ04584L, UDI/DI 10198459212291 (each) 40198459212292 (case), Lot number 26DMA968; Medline product number/SKU DYNJT5715A, UDI/DI 10198459505126 (each) 40198459505127 (case), Lot number 26CMI506; Medline product number/SKU DYNJ85751, UDI/DI 10195327542917 (each) 40195327542918 (case), Lot number 26BMB785; Medline product number/SKU DYNJ53741N, UDI/DI 10198459560088 (each) 40198459560089 (case), Lot number 26BBG793; Medline product number/SKU DYNJ80701A, UDI/DI 10198459026195 (each) 40198459026196 (case), Lot number 26BBF945; Medline product number/SKU DYNJ67308C, UDI/DI 10198459676666 (each) 40198459676667 (case), Lot number 26BBF661; Medline product number/SKU DYNJ89451B, UDI/DI 10198459637674 (each) 40198459637675 (case), Lot number 26BBC903; Medline product number/SKU DYNJ42166A, UDI/DI 10888277938281 (each) 40888277938282 (case), Lot number 26BBB320; Medline product number/SKU DYNJ67363C, UDI/DI 10198459652226 (each) 40198459652227 (case), Lot number 26AML066; Medline product number/SKU DYNJ67363B, UDI/DI 10198459259081 (each) 40198459259082 (case), Lot number 26AMK067; Medline product number/SKU DYNJ65779B, UDI/DI 10195327571733 (each) 40195327571734 (case), Lot number 26AMI446; Medline product number/SKU DYNJ47661J, UDI/DI 10198459525704 (each) 40198459525705 (case), Lot number 26AMF686; Medline product number/SKU DYNJ04584L, UDI/DI 10198459212291 (each) 40198459212292 (case), Lot number 26AMF245; Medline product number/SKU DYNJ04470L, UDI/DI
Distribution
US Nationwide distribution.
What does Class II mean?
Class II means use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or the probability of serious harm is remote.
Source: FDA (medical devices), recall Z-0176-2027. Official notice. Republished from public data; RecallFlag is not affiliated with the FDA.