FDA Medical devices recall ·

Medline surgical convenience kits labeled as: OPEN HEART… recall

Recall number
Z-0178-2027
Date
(initiated September 2, 2026)
Company / brand
Medline Industries, LP
3 Lakes Dr, Northfield, IL, United States
Category
Medical devices › Class II
Classification
Class II
Status
Ongoing
Units affected
165 kits
Initiated by
Voluntary: Firm initiated

Read the official FDA notice ↗

Product

Medline surgical convenience kits labeled as: OPEN HEART, Medline Product Number/SKU DYNJ902898D

Why was it recalled?

Medical convenience kits were distributed containing cannula components which do not conform to ISO 80369-7 dimensional requirements and may contribute to leaking or breaking during use.

What should you do?

FDA enforcement reports do not list a consumer remedy. Check the product codes below against what you have; if it matches, stop using it and follow the instructions in the firm's or FDA's official notice, or contact the company. If you have health concerns, contact your healthcare provider.

Product codes & lot numbers

UDI/DI 10195327369743 (each) 40195327369744 (case). Lot Numbers: 23GBD734, 23EBQ512

Distribution

US Nationwide distribution.

What does Class II mean?

Class II means use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or the probability of serious harm is remote.

Source: FDA (medical devices), recall Z-0178-2027. Official notice. Republished from public data; RecallFlag is not affiliated with the FDA.