FDA Medical devices recall ·
TS Misc. Instruments Tray, Triathlon. Catalog Number: 6543-6-519 recall
- Recall number
- Z-0167-2027
- Date
- (initiated September 16, 2026)
- Company / brand
- Howmedica Osteonics Corp.
325 Corporate Dr, Mahwah, NJ, United States - Category
- Medical devices › Class II
- Classification
- Class II
- Status
- Ongoing
- Units affected
- 544 trays
- Hazard tags
- Sterility
- Initiated by
- Voluntary: Firm initiated
Read the official FDA notice ↗
Product
TS Misc. Instruments Tray, Triathlon. Catalog Number: 6543-6-519.
Why was it recalled?
Reprocessing instructions provided in LSTPI-B specify a sterilization dry time of 30 minutes for the trays sterilized in blue wrap. When blue wrap is used, the 30-minute dry time may lead to a small amount of residual moisture remaining after sterilization.
What should you do?
FDA enforcement reports do not list a consumer remedy. Check the product codes below against what you have; if it matches, stop using it and follow the instructions in the firm's or FDA's official notice, or contact the company. If you have health concerns, contact your healthcare provider.
Product codes & lot numbers
Catalog Number: 6543-6-519. GTIN: 07613327555127. All lot numbers used prior to the correction date of May 8, 2026 are impacted.
Distribution
US Nationwide distribution. International distribution to Canada, Australia, New Zealand.
What does Class II mean?
Class II means use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or the probability of serious harm is remote.
Source: FDA (medical devices), recall Z-0167-2027. Official notice. Republished from public data; RecallFlag is not affiliated with the FDA.