Triathlon Hinge Femoral Prep Tray recall

Recall number
Z-0166-2027
Date
(initiated September 16, 2026)
Company / brand
Howmedica Osteonics Corp.
325 Corporate Dr, Mahwah, NJ, United States
Category
Medical devices › Class II
Classification
Class II
Status
Ongoing
Units affected
340 trays
Hazard tags
Sterility
Initiated by
Voluntary: Firm initiated

Read the official FDA notice ↗

Product

Triathlon Hinge Femoral Prep Tray. Catalog Number: 6543-6-514.

Why was it recalled?

Reprocessing instructions provided in LSTPI-B specify a sterilization dry time of 30 minutes for the trays sterilized in blue wrap. When blue wrap is used, the 30-minute dry time may lead to a small amount of residual moisture remaining after sterilization.

What should you do?

FDA enforcement reports do not list a consumer remedy. Check the product codes below against what you have; if it matches, stop using it and follow the instructions in the firm's or FDA's official notice, or contact the company. If you have health concerns, contact your healthcare provider.

Product codes & lot numbers

Catalog Number: 6543-6-514. GTIN: 07613327562859. All lot numbers used prior to the correction date of May 8, 2026 are impacted.

Distribution

US Nationwide distribution. International distribution to Canada, Australia, New Zealand.

What does Class II mean?

Class II means use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or the probability of serious harm is remote.

Source: FDA (medical devices), recall Z-0166-2027. Official notice. Republished from public data; RecallFlag is not affiliated with the FDA.