FDA Medical devices recall ·

Presource Packs Containing Stryker StrykeFlow PS and AHTO tube sets: 1)… recall

Recall number
Z-3298-2026
Date
(initiated August 18, 2026)
Company / brand
Cardinal Health 200, LLC
3651 Birchwood Dr, Waukegan, IL, United States
Category
Medical devices › Class II
Classification
Class II
Status
Ongoing
Units affected
22 kits
Hazard tags
Fall
Initiated by
Voluntary: Firm initiated

Read the official FDA notice ↗

Product

Presource Packs Containing Stryker StrykeFlow PS and AHTO tube sets: 1) KIT, LAVH , Catalog Number: PB63VHSTC

Why was it recalled?

The kits contain tube sets with handpieces that may leak during use. The impacted devices can have the following failures: fluid leakage from the hand piece that can pool on the floor. The impact can range from user dissatisfaction to the potential for the user of slip and fall.

What should you do?

FDA enforcement reports do not list a consumer remedy. Check the product codes below against what you have; if it matches, stop using it and follow the instructions in the firm's or FDA's official notice, or contact the company. If you have health concerns, contact your healthcare provider.

Product codes & lot numbers

1) PB63VHSTC, UDI-DI: 10197106980866(each), 50197106980864(case), Lot Number: 25069FG2 ; 2) PB63VHSTC, UDI-DI: 10197106980866(each), 50197106980864(case), Lot Number: 25101FG2

Distribution

US Nationwide distribution in the states of CA, IL, KY, MO, NC, NY, OH, TX.

States: California Illinois Kentucky Missouri North Carolina New York Ohio Texas

What does Class II mean?

Class II means use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or the probability of serious harm is remote.

Source: FDA (medical devices), recall Z-3298-2026. Official notice. Republished from public data; RecallFlag is not affiliated with the FDA.