FDA Medical devices recall ·
Presource Packs Containing Stryker StrykeFlow PS and AHTO tube sets: 1)… recall
- Recall number
- Z-3296-2026
- Date
- (initiated August 18, 2026)
- Company / brand
- Cardinal Health 200, LLC
3651 Birchwood Dr, Waukegan, IL, United States - Category
- Medical devices › Class II
- Classification
- Class II
- Status
- Ongoing
- Units affected
- 344 kits
- Hazard tags
- Fall
- Initiated by
- Voluntary: Firm initiated
Read the official FDA notice ↗
Product
Presource Packs Containing Stryker StrykeFlow PS and AHTO tube sets: 1) KIT,GU DAVINCI , Catalog Number: PGVMDVBWU; 2) KIT,GU DAVINCI , Catalog Number: PGVMDVBWV; 3) KIT,GU DAVINCI , Catalog Number: PGVMDVB26; 4) KIT,GU DAVINCI , Catalog Number: PGVMDVB27; 5) KIT,GU DAVINCI , Catalog Number: PGVMDVB28; 6) KIT,GU DAVINCI , Catalog Number: PGVMDVB29; 7) COLORECTAL PACK , Catalog Number: SBA73CLKFD; 8) COLORECTAL PACK , Catalog Number: SBA73CLK26
Why was it recalled?
The kits contain tube sets with handpieces that may leak during use. The impacted devices can have the following failures: fluid leakage from the hand piece that can pool on the floor. The impact can range from user dissatisfaction to the potential for the user of slip and fall.
What should you do?
FDA enforcement reports do not list a consumer remedy. Check the product codes below against what you have; if it matches, stop using it and follow the instructions in the firm's or FDA's official notice, or contact the company. If you have health concerns, contact your healthcare provider.
Product codes & lot numbers
1) PGVMDVBWU, UDI-DI: 10888439626186(each), 50888439626184(case), Lot Number: 24317FG2 ; 2) PGVMDVBWU, UDI-DI: 10888439626186(each), 50888439626184(case), Lot Number: 24353FG2 ; 3) PGVMDVBWU, UDI-DI: 10888439626186(each), 50888439626184(case), Lot Number: 24317FG2 ; 4) PGVMDVBWU, UDI-DI: 10888439626186(each), 50888439626184(case), Lot Number: 25069FG2 ; 5) PGVMDVBWU, UDI-DI: 10888439626186(each), 50888439626184(case), Lot Number: 25045FG2 ; 6) PGVMDVBWU, UDI-DI: 10888439626186(each), 50888439626184(case), Lot Number: 25069FG2 ; 7) PGVMDVBWV, UDI-DI: 10198956204034(each), 50198956204032(case), Lot Number: 24353FG2 ; 8) PGVMDVBWV, UDI-DI: 10198956204034(each), 50198956204032(case), Lot Number: 25101FG2 ; 9) PGVMDVBWV, UDI-DI: 10198956204034(each), 50198956204032(case), Lot Number: 25148FG2 ; 10) PGVMDVBWV, UDI-DI: 10198956204034(each), 50198956204032(case), Lot Number: 25217FG2 ; 11) PGVMDVB26, UDI-DI: 10198956007796(each), 50198956007794(case), Lot Number: 24309FG2 ; 12) PGVMDVB27, UDI-DI: 10198956054714(each), 50198956054712(case), Lot Number: 24325FG2 ; 13) PGVMDVB28, UDI-DI: 10198956079847(each), 50198956079845(case), Lot Number: 24317FG2 ; 14) PGVMDVB28, UDI-DI: 10198956079847(each), 50198956079845(case), Lot Number: 24325FG2 ; 15) PGVMDVB29, UDI-DI: 10198956094024(each), 50198956094022(case), Lot Number: 24317FG2 ; 16) PGVMDVB29, UDI-DI: 10198956094024(each), 50198956094022(case), Lot Number: 24348FG2 ; 17) SBA73CLKFD, UDI-DI: 10197106633960(each), 50197106633968(case), L
Distribution
US Nationwide distribution in the states of CA, IL, KY, MO, NC, NY, OH, TX.
States: California Illinois Kentucky Missouri North Carolina New York Ohio Texas
What does Class II mean?
Class II means use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or the probability of serious harm is remote.
Source: FDA (medical devices), recall Z-3296-2026. Official notice. Republished from public data; RecallFlag is not affiliated with the FDA.