FDA Medical devices recall ·

Medline breathing circuits labeled as: 1) MEDLINE HUDSON RCI, CIRCUIT… recall

Recall number
Z-2993-2026
Date
(initiated July 13, 2026)
Company / brand
Medline Industries, LP
3 Lakes Dr, Northfield, IL, United States
Category
Medical devices › Class II
Classification
Class II
Status
Ongoing
Units affected
271406 units
Hazard tags
Burn
Initiated by
Voluntary: Firm initiated

Read the official FDA notice ↗

Product

Medline breathing circuits labeled as: 1) MEDLINE HUDSON RCI, CIRCUIT, NEO, SING LIMB, HTD,COLUMN, Medline Product Number/SKU HUD78001KIT (Teleflex Product Number/SKU 780-01KIT); 2) MEDLINE HUDSON RCI, CIRCUIT, NEO, DUAL LIMB HTD, COLUMN, Medline Product Number/SKU HUD78007KIT (Teleflex Product Number/SKU 780-07KIT); 3) MEDLINE HUDSON RCI, DBD-CIRCUIT, NEO, DUAL LIMB HTD, COLUMN, Medline Product Number/SKU HUD78008KIT (Teleflex Product Number/SKU 780-08KIT); 4) MEDLINE HUDSON RCI, DBD-CIRCUIT, NEO, DUAL LIMB HTD, COLUMN, Medline Product Number/SKU HUD78009KIT (Teleflex Product Number/SKU 790-09KIT); 5) MEDLINE HUDSON RCI, CIRCUIT, NEO, DUAL LIMB, HTD,COLUMN, Medline Product Number/SKU HUD78017KIT (Teleflex Product Number/SKU 780-17KIT); 6) MEDLINE HUDSON RCI, CIRCUIT, NEO, DUAL HTD, DUAL LIMB,COLUMN, Medline Product Number/SKU HUD78018KIT (Teleflex Product Number/SKU 780-18KIT); 7) MEDLINE HUDSON RCI, CIRCUIT,ISO-SHIELD SINGLE LIMB,INFANT, Medline Product Number/SKU HUD88001KIT (Teleflex Product Number/SKU 880-01KIT); 8) MEDLINE HUDSON RCI, CIRCUIT,ISO-SHIELD,INFANT,HTD,COLUMN,DRN, Medline Product Number/SKU HUD88013KIT (Teleflex Product Number/SKU 880-13KIT); 9) MEDLINE HUDSON RCI, ISO-SHIELD CIRCUIT,INFNT,HTD,DL,DRAIN, Medline Product Number/SKU HUD88014KIT (Teleflex Product Number/SKU 880-14KIT); 10) MEDLINE HUDSON RCI,ISO-SHIELD CIRCUIT,INFNT,HTD,DL,DRAIN, Medline Product Number/SKU HUD88015KIT (Teleflex Product Number/SKU 880-15KIT); 11) MEDLINE HUDSON RCI, ISO-SHIELD CIRCUIT,INFNT,HTD,DL,DRAIN, Medline Product Number/SKU HUD88017KIT (Teleflex Product Number/SKU 880-17KIT).

Why was it recalled?

Neonatal and infant heated wire breathing circuits may experience sporadic failure. The heated wire connector (pigtail) at the humidifier may exhibit discoloration, melting, smoke, burning odor, or other signs of localized thermal damage. Device failure may predispose infant to bronchospasm, inspissated secretions, and airway obstruction, as well as inadvertent positive end-expiratory pressure (PEEP), circuit obstruction, or accidental lavage of pooled fluid into the airway, elevated oxygen consumption, and worsening respiratory distress.

What should you do?

FDA enforcement reports do not list a consumer remedy. Check the product codes below against what you have; if it matches, stop using it and follow the instructions in the firm's or FDA's official notice, or contact the company. If you have health concerns, contact your healthcare provider.

Product codes & lot numbers

1) Medline Product Number/SKU HUD78001KIT (Teleflex Product Number/SKU 780-01KIT), UDI/DI 10197344010752 (each) 20197344010759 (case), All lot numbers; 2) Medline Product Number/SKU HUD78007KIT (Teleflex Product Number/SKU 780-07KIT), UDI/DI 10197344010738 (each) 20197344010735 (case), All lot numbers; 3) Medline Product Number/SKU HUD78008KIT (Teleflex Product Number/SKU 780-08KIT), UDI/DI 10197344010721 (each) 20197344010728 (case), All lot numbers; 4) Medline Product Number/SKU HUD78009KIT (Teleflex Product Number/SKU 790-09KIT), UDI/DI 10197344010714 (each) 20197344010711 (case), All lot numbers; 5) Medline Product Number/SKU HUD78017KIT (Teleflex Product Number/SKU 780-17KIT), UDI/DI 10197344010677 (each) 20197344010674 (case), All lot numbers; 6) Medline Product Number/SKU HUD78018KIT (Teleflex Product Number/SKU 780-18KIT), UDI/DI 10197344010905 (each) 20197344010902 (case), All lot numbers; 7) Medline Product Number/SKU HUD88001KIT (Teleflex Product Number/SKU 880-01KIT), UDI/DI 10197344010929 (each) 20197344010926 (case), All lot numbers; 8) Medline Product Number/SKU HUD88013KIT (Teleflex Product Number/SKU 880-13KIT), UDI/DI 10197344010493 (each) 20197344010490 (case), All lot numbers; 9) Medline Product Number/SKU HUD88014KIT (Teleflex Product Number/SKU 880-14KIT), UDI/DI 10197344010486 (each) 20197344010483 (case), All lot numbers; 10) Medline Product Number/SKU HUD88015KIT (Teleflex Product Number/SKU 880-15KIT), UDI/DI 10197344010479 (each) 20197344010476 (cas

Distribution

Worldwide distribution - US Nationwide and the country of Panama.

What does Class II mean?

Class II means use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or the probability of serious harm is remote.

Source: FDA (medical devices), recall Z-2993-2026. Official notice. Republished from public data; RecallFlag is not affiliated with the FDA.