FDA Medical devices recall ·

Medline breathing circuits labeled as: 1) HUDSON RCI, CIRCUIT, HTD, DL… recall

Recall number
Z-2992-2026
Date
(initiated July 13, 2026)
Company / brand
Medline Industries, LP
3 Lakes Dr, Northfield, IL, United States
Category
Medical devices › Class II
Classification
Class II
Status
Ongoing
Units affected
145525 units
Hazard tags
Burn
Initiated by
Voluntary: Firm initiated

Read the official FDA notice ↗

Product

Medline breathing circuits labeled as: 1) HUDSON RCI, CIRCUIT, HTD, DL, NEONATE, 48, Medline Product Number/SKU HUD78007 (Teleflex Product Number/SKU 780-07); 2) HUDSON RCI, DBM-CIRCUIT,VENT,NEONATAL,21 V, Medline Product Number/SKU HUD78009 (Teleflex Product Number/SKU 780-09); 3) HUDSON RCI, DBD-CIRCUIT,VENT,INSIRATORY,HEATED,NEO,4, Medline Product Number/SKU 780-13 (Teleflex Product Number/SKU 780-13); 4) HUDSON RCI, DBD-CIRCUIT,VENT,INSIRATORY,HEATED,NEO,4, Medline Product Number/SKU 780-14 (Teleflex Product Number/SKU 780-14); 5) HUDSON RCI, DBD-CIRCUIT, HTD, DUAL LIMB, NEONATE, Medline Product Number/SKU 780-15 (Teleflex Product Number/SKU 780-15); 6) HUDSON RCI, CIRCUIT,NEO,DUAL,WIRE,60 IN, Medline Product Number/SKU HUD78018) (Teleflex Product Number/SKU 780-18).

Why was it recalled?

Neonatal and infant heated wire breathing circuits may experience sporadic failure. The heated wire connector (pigtail) at the humidifier may exhibit discoloration, melting, smoke, burning odor, or other signs of localized thermal damage. Device failure may predispose infant to bronchospasm, inspissated secretions, and airway obstruction, as well as inadvertent positive end-expiratory pressure (PEEP), circuit obstruction, or accidental lavage of pooled fluid into the airway, elevated oxygen consumption, and worsening respiratory distress.

What should you do?

FDA enforcement reports do not list a consumer remedy. Check the product codes below against what you have; if it matches, stop using it and follow the instructions in the firm's or FDA's official notice, or contact the company. If you have health concerns, contact your healthcare provider.

Product codes & lot numbers

1) Medline Product Number/SKU HUD78007, UDI/DI 10197344042463 (each) 20197344042460 (case), All lot numbers; 2) Medline Product Number/SKU HUD78009, UDI/DI Discontinued (each) Discontinued (case), All lot numbers; 3) Medline Product Number/SKU 780-13, UDI/DI Discontinued (each) Discontinued (case), All lot numbers; 4) Medline Product Number/SKU 780-14, UDI/DI Discontinued (each) Discontinued (case), All lot numbers; 5) Medline Product Number/SKU 780-15, UDI/DI Discontinued (each) Discontinued (case), All lot numbers; 6) Medline Product Number/SKU HUD78018, UDI/DI 10197344010660 (each) 20197344010667 (case), All lot numbers.

Distribution

Worldwide distribution - US Nationwide and the country of Panama.

What does Class II mean?

Class II means use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or the probability of serious harm is remote.

Source: FDA (medical devices), recall Z-2992-2026. Official notice. Republished from public data; RecallFlag is not affiliated with the FDA.