FDA Medical devices recall ·

Reprocessed Inquiry Steerable Diagnostics Catheter for electrogram… recall

This is an FDA Class I recall — the most serious type.

Recall number
Z-2612-2025
Date
(initiated July 31, 2025)
Company / brand
Medline Industries, LP
3 Lakes Dr, Northfield, IL, United States
Category
Medical devices › Class I
Classification
Class I
Status
Ongoing
Units affected
7913 eaches
Initiated by
Voluntary: Firm initiated

Read the official FDA notice ↗

Product

Reprocessed Inquiry Steerable Diagnostics Catheter for electrogram recording and cardiac stimulation during diagnostic electrophysiologic studies; Medline Item Numbers 81102RH 81104RH 81107RH 81174RH 81402RH 81404RH 81405RH 81472RH 81473RH 81474RH 81531RH 81532RH

Why was it recalled?

Specific lots of reprocessed electrophysiology and ultrasound catheters may contain small residual particulates. If devices with residual particulates on patient-contacting surfaces are used, there is a potential risk of an inflammatory response or systemic infection, and if particulate were to dislodge during use and contact circulating blood and/or the heart, it may produce a granulomatous reaction or thrombus formation, which may lead to cerebral or pulmonary embolism, or deep vein thrombosis.

What should you do?

FDA enforcement reports do not list a consumer remedy. Check the product codes below against what you have; if it matches, stop using it and follow the instructions in the firm's or FDA's official notice, or contact the company. If you have health concerns, contact your healthcare provider.

Product codes & lot numbers

Item Number/UDI-DI 81102RH 10197344133239; 81104RH 10197344133246; 81107RH 10197344133260; 81174RH 10197344133291; 81402RH 10197344133352; 81404RH 10197344133376; 81405RH 10197344133383; 81472RH 10197344133413; 81473RH 10197344133420; 81474RH 10197344133437; 81531RH 10197344133512; 81532RH 10197344133529; LOTS EP250212 EP250214

Distribution

Domestic US distribution nationwide. International distribution pending. No international distribution.

What does Class I mean?

Class I is the most serious FDA recall class: there is a reasonable probability that using or being exposed to the product will cause serious adverse health consequences or death.

Source: FDA (medical devices), recall Z-2612-2025. Official notice. Republished from public data; RecallFlag is not affiliated with the FDA.