FDA Medical devices recall ·
Medline Convenience kits containing recalled Swan-Ganz Catheters RIGHT… recall
- Recall number
- Z-2606-2026
- Date
- (initiated May 28, 2026)
- Company / brand
- Medline Industries, LP
3 Lakes Dr, Northfield, IL, United States - Category
- Medical devices › Class II
- Classification
- Class II
- Status
- Ongoing
- Units affected
- 9,272 kits
- Initiated by
- Voluntary: Firm initiated
Read the official FDA notice ↗
Product
Medline Convenience kits containing recalled Swan-Ganz Catheters RIGHT HEART CATH PACK CTX DYNJ68530B
Why was it recalled?
Kits contain Swan-Ganz catheters recalled by Becton Dickinson (BD). BD has received complaints of leaks/breakage involving failures of the (blue) proximal injectate lumen hub on these catheters. This has led to instances of leakage, lumen damage, and breakage, and may lead to infection, medication loss, and/or blood loss.
What should you do?
FDA enforcement reports do not list a consumer remedy. Check the product codes below against what you have; if it matches, stop using it and follow the instructions in the firm's or FDA's official notice, or contact the company. If you have health concerns, contact your healthcare provider.
Product codes & lot numbers
UDI-DI 10195327345044 Lots 23DBI807 23FBG708 23GBK642 23KBO477 24BBG198 24FBN889 24GBJ879 24HBR045 24LBT492 25ABL957 25BBN572 25GBT734 25HBQ817 25LBM617 26BBK766 26DBE720
Distribution
US Nationwide distribution in the states of AZ, CA, FL, IL, MO, NJ, OH, OK, TN, TX, WI.
States: Arizona California Florida Illinois Missouri New Jersey Ohio Oklahoma Tennessee Texas Wisconsin
What does Class II mean?
Class II means use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or the probability of serious harm is remote.
Source: FDA (medical devices), recall Z-2606-2026. Official notice. Republished from public data; RecallFlag is not affiliated with the FDA.