FDA Medical devices recall ·

Medline Convenience kits containing recalled Swan-Ganz Catheters RIGHT… recall

Recall number
Z-2612-2026
Date
(initiated May 28, 2026)
Company / brand
Medline Industries, LP
3 Lakes Dr, Northfield, IL, United States
Category
Medical devices › Class II
Classification
Class II
Status
Ongoing
Units affected
9,272 kits
Initiated by
Voluntary: Firm initiated

Read the official FDA notice ↗

Product

Medline Convenience kits containing recalled Swan-Ganz Catheters RIGHT HEART CATH PACK CTX DYNJ68530A

Why was it recalled?

Kits contain Swan-Ganz catheters recalled by Becton Dickinson (BD). BD has received complaints of leaks/breakage involving failures of the (blue) proximal injectate lumen hub on these catheters. This has led to instances of leakage, lumen damage, and breakage, and may lead to infection, medication loss, and/or blood loss.

What should you do?

FDA enforcement reports do not list a consumer remedy. Check the product codes below against what you have; if it matches, stop using it and follow the instructions in the firm's or FDA's official notice, or contact the company. If you have health concerns, contact your healthcare provider.

Product codes & lot numbers

UDI-DI 10195327033620 Lots 21KBN891 21KBO410 22ABA676 22ABY937 22BBN909 22BBT447 23ABO303

Distribution

US Nationwide distribution in the states of AZ, CA, FL, IL, MO, NJ, OH, OK, TN, TX, WI.

States: Arizona California Florida Illinois Missouri New Jersey Ohio Oklahoma Tennessee Texas Wisconsin

What does Class II mean?

Class II means use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or the probability of serious harm is remote.

Source: FDA (medical devices), recall Z-2612-2026. Official notice. Republished from public data; RecallFlag is not affiliated with the FDA.