FDA Medical devices recall ·
MEDLINE Medical Procedure Kits labeled as: ADULT TRACHEOSTOMY TUBE INSERT… recall
- Recall number
- Z-2491-2026
- Date
- (initiated May 4, 2026)
- Company / brand
- Medline Industries, LP
3 Lakes Dr, Northfield, IL, United States - Category
- Medical devices › Class II
- Classification
- Class II
- Status
- Ongoing
- Units affected
- 217390 units
- Initiated by
- Voluntary: Firm initiated
Read the official FDA notice ↗
Product
MEDLINE Medical Procedure Kits labeled as: ADULT TRACHEOSTOMY TUBE INSERT, ASC T&A, CONWAY T&A SINUS, EAR PACK, etc. (see the recall documents for the full list of products)
Why was it recalled?
During an internal review, Medline identified representative samples did not meet applicable performance specifications. Identified issues include irregularities in catheter surface finish, instances where the catheter connector detached prematurely from the catheter funnel, and limited cases where the catheter outer diameter did not meet specification.
What should you do?
FDA enforcement reports do not list a consumer remedy. Check the product codes below against what you have; if it matches, stop using it and follow the instructions in the firm's or FDA's official notice, or contact the company. If you have health concerns, contact your healthcare provider.
Product codes & lot numbers
Medline Kit Number/SKU MNS12180: UDI/DI 10653160327134 (each), 40653160327135 (case), Lot Number 22FBM309; Medline Kit Number/SKU MNS12180: UDI/DI 10653160327134 (each), 40653160327135 (case), Lot Number 22DBJ094; Medline Kit Number/SKU MNS12180: UDI/DI 10653160327134 (each), 40653160327135 (case), Lot Number 22OBH091; Medline Kit Number/SKU MNS12180: UDI/DI 10653160327134 (each), 40653160327135 (case), Lot Number 21HBC428; Medline Kit Number/SKU DYNJ67109B: UDI/DI 10195327207083 (each), 40195327207084 (case), Lot Number 24ABC426; Medline Kit Number/SKU DYNJ67109B: UDI/DI 10195327207083 (each), 40195327207084 (case), Lot Number 23JBM836; Medline Kit Number/SKU DYNJ67109B: UDI/DI 10195327207083 (each), 40195327207084 (case), Lot Number 23HBS357; Medline Kit Number/SKU DYNJ67109B: UDI/DI 10195327207083 (each), 40195327207084 (case), Lot Number 22KBS784; Medline Kit Number/SKU DYNJ67109B: UDI/DI 10195327207083 (each), 40195327207084 (case), Lot Number 22IBV574; Medline Kit Number/SKU DYNJ67109D: UDI/DI 10195327689100 (each), 40195327689101 (case), Lot Number 26ABH211; Medline Kit Number/SKU DYNJ67109D: UDI/DI 10195327689100 (each), 40195327689101 (case), Lot Number 25GBS702; Medline Kit Number/SKU DYNJ67109D: UDI/DI 10195327689100 (each), 40195327689101 (case), Lot Number 25EBT418; Medline Kit Number/SKU DYNJ67109D: UDI/DI 10195327689100 (each), 40195327689101 (case), Lot Number 24LBP073; Medline Kit Number/SKU DYNJ67109D: UDI/DI 10195327689100 (each), 40195327689101 (case), Lot
Distribution
Worldwide distribution - US Nationwide and the countries of AE, BB, BM, BS, CA, CY, GU, NL, PA, QA, SA, VI.
States: California Pennsylvania
What does Class II mean?
Class II means use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or the probability of serious harm is remote.
Source: FDA (medical devices), recall Z-2491-2026. Official notice. Republished from public data; RecallFlag is not affiliated with the FDA.