FDA Medical devices recall ·

MEDLINE Medical Procedure Kits labeled as: 1) BNT CV PACK… recall

Recall number
Z-2488-2026
Date
(initiated May 4, 2026)
Company / brand
Medline Industries, LP
3 Lakes Dr, Northfield, IL, United States
Category
Medical devices › Class II
Classification
Class II
Status
Ongoing
Units affected
1248 units
Initiated by
Voluntary: Firm initiated

Read the official FDA notice ↗

Product

MEDLINE Medical Procedure Kits labeled as: 1) BNT CV PACK, Medline Kit Number/SKU DYNJ912271; 2) CV PACK PART 1, Medline Kit Number/SKU DYNJ31204D; 3) KIT PERIPHERAL VASCULAR MMC, Medline Kit Number/SKU DYNJ905368D; 4) KIT PERIVASCULAR MMC, Medline Kit Number/SKU DYNJ905368B; 5) VP SHUNT PACK, Medline Kit Number/SKU DYNJ46530I.

Why was it recalled?

During an internal review, Medline identified representative samples did not meet applicable performance specifications. Identified issues include irregularities in catheter surface finish, instances where the catheter connector detached prematurely from the catheter funnel, and limited cases where the catheter outer diameter did not meet specification.

What should you do?

FDA enforcement reports do not list a consumer remedy. Check the product codes below against what you have; if it matches, stop using it and follow the instructions in the firm's or FDA's official notice, or contact the company. If you have health concerns, contact your healthcare provider.

Product codes & lot numbers

Medline Kit Number/SKU DYNJ912271: UDI/DI 10198459688324 (each), 40198459688325 (case), Lot Number 26CBQ109; Medline Kit Number/SKU DYNJ31204D: UDI/DI 10193489957990 (each), 40193489957991 (case), Lot Number 25FBO720; Medline Kit Number/SKU DYNJ31204D: UDI/DI 10193489957990 (each), 40193489957991 (case), Lot Number 24KBO387; Medline Kit Number/SKU DYNJ31204D: UDI/DI 10193489957990 (each), 40193489957991 (case), Lot Number 24KBO015; Medline Kit Number/SKU DYNJ31204D: UDI/DI 10193489957990 (each), 40193489957991 (case), Lot Number 24KBD819; Medline Kit Number/SKU DYNJ31204D: UDI/DI 10193489957990 (each), 40193489957991 (case), Lot Number 24KBA736; Medline Kit Number/SKU DYNJ31204D: UDI/DI 10193489957990 (each), 40193489957991 (case), Lot Number 24FBG144; Medline Kit Number/SKU DYNJ31204D: UDI/DI 10193489957990 (each), 40193489957991 (case), Lot Number 24DBO838; Medline Kit Number/SKU DYNJ31204D: UDI/DI 10193489957990 (each), 40193489957991 (case), Lot Number 24BBC453; Medline Kit Number/SKU DYNJ31204D: UDI/DI 10193489957990 (each), 40193489957991 (case), Lot Number 24ABM118; Medline Kit Number/SKU DYNJ31204D: UDI/DI 10193489957990 (each), 40193489957991 (case), Lot Number 23KBS950; Medline Kit Number/SKU DYNJ31204D: UDI/DI 10193489957990 (each), 40193489957991 (case), Lot Number 23IBC670; Medline Kit Number/SKU DYNJ31204D: UDI/DI 10193489957990 (each), 40193489957991 (case), Lot Number 22LBN883; Medline Kit Number/SKU DYNJ31204D: UDI/DI 10193489957990 (each), 40193489957991 (ca

Distribution

Worldwide distribution - US Nationwide and the countries of AE, BB, BM, BS, CA, CY, GU, NL, PA, QA, SA, VI.

States: California Pennsylvania

What does Class II mean?

Class II means use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or the probability of serious harm is remote.

Source: FDA (medical devices), recall Z-2488-2026. Official notice. Republished from public data; RecallFlag is not affiliated with the FDA.