FDA Medical devices recall ·
Giraffe OmniBed, a combination of an infant incubator and an infant… recall
- Recall number
- Z-2475-2026
- Date
- (initiated May 1, 2026)
- Company / brand
- GE Medical Systems, LLC
3000 N Grandview Blvd, Waukesha, WI, United States - Category
- Medical devices › Class II
- Classification
- Class II
- Status
- Ongoing
- Units affected
- 19587 units
- Hazard tags
- Fall Injury risk
- Initiated by
- Voluntary: Firm initiated
Read the official FDA notice ↗
Product
Giraffe OmniBed, a combination of an infant incubator and an infant warmer, All Models
Why was it recalled?
GE HealthCare has become aware of a potential issue involving the canopy soffit beneath the control panel end of the device. Under certain conditions, for example, impact during transport, one or more of the fasteners that secure the soffit can disengage. If this occurs, a fastener could fall into the patient compartment. In rare circumstances, if multiple fasteners disengage, the canopy soffit can become loose or detach. These conditions could potentially result in patient injury.
What should you do?
FDA enforcement reports do not list a consumer remedy. Check the product codes below against what you have; if it matches, stop using it and follow the instructions in the firm's or FDA's official notice, or contact the company. If you have health concerns, contact your healthcare provider.
Product codes & lot numbers
Serial Numbers: HDGP50034 HDGP51018 HDGS50069 HDGS50070 HDGS50074 HDGS50077 HDGS50158 HDGS50162 HDGS52245 HDGS52286 HDGT50374 HDGT50399 HDGS50464 HDGS50560 HDGS50569 HDGS50570 HDGD50349 HDGG54079 HDGJ50278 HDGJ50281 HDGJ50390 HDGJ50391 HDGJ50394 HDGJ50400 HDGK52363 HDGM50481 HDGM50482 HDGM50483 HDGM50484 HDGM50485 HDGM50486 HDGQ50625 HDGQ50626 HDGQ50627 HDGQ50628 HDGQ50629 HDGS50369 HDGS50370 HDGS50371 HDGS50372 HDGR50086 HDGL51597 HDGG54268 HDGG54270 HDGG54410 HDGK51696 HDGL51002 HDGL51003 HDGL51004 HDGL51005 HDGL51032 HDGN50383 HDGP50522 HDGQ50763 HDGQ50882 HDGQ50884 HDGQ50886 HDGQ50887 HDGQ50901 HDGQ50902 HDGQ50903 HDGQ50904 HDGQ50905 HDGQ50906 HDGQ50907 HDGQ50908 HDGQ50909 HDGQ50910 HDGQ50911 HDGQ50912 HDGQ50913 HDGQ50917 HDGS51833 HDGU60609 HDGU60732 HDGE50297 HDGE50370 HDGE50371 HDGK51791 HDGK51792 HDGK51793 HDGK51794 HDGK51795 HDGK51804 HDGK51822 HDGK51840 HDGR50087 HDGR50088 HDGR50178 HDGR50180 HDGR50184 HDGR50185 HDGN50863 HDGN50864 HDGN50865 HDGN50866 HDGN50868 HDGN50869 HDGN50870 HDGH60335 HDGH60316 HDGH60319 HDGH60320 HDGH60321 HDGH60322 HDGH60323 HDGH60324 HDGH60326 HDGH60327 HDGH60328 HDGH60329 HDGH60330 HDGH60331 HDGH60333 HDGH60334 HDGH60403 HDGH60407 HDGH60408 HDGH60409 HDGN50867 HDGT50420 HDGK51412 HDGK51706 HDGK52164 HDGL50039 HDGL51294 HDGL51483 HDGL51486 HDGM51574 HDGM51575 HDGM51583 HDGP51029 HDGS52343 HDGS52358 HDGJ50055 HDGN50393 HDGM50726 HDGM50093 HDGM50146 HDGM50149 HDGM50151 HDGP51237 HDGP51290 HDGP51297 HDGQ51823 HDGQ51947 HDGJ51132 HDGJ51133 HDGJ
Distribution
Worldwide - US Nationwide distribution.
What does Class II mean?
Class II means use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or the probability of serious harm is remote.
Source: FDA (medical devices), recall Z-2475-2026. Official notice. Republished from public data; RecallFlag is not affiliated with the FDA.